Adaptive dynamic range matching for spectroscopic measurements.

Published

Journal Article

We describe the design and development of an adaptive spectrometer that actively manages dynamic range mismatch between source signals and the spectrometer's detector. The introduction of an inexpensive digital micromirror array yields performance improvements compared to a traditional device. Simulation results are presented that indicate real-world scenarios in which the adaptive spectrometer will yield superior performance. Experimental results from a first-generation prototype adaptive spectrometer confirm the expectations of the simulations.

Full Text

Duke Authors

Cited Authors

  • Kinast, J; Gehm, ME

Published Date

  • April 2009

Published In

Volume / Issue

  • 48 / 10

Start / End Page

  • 1891 - 1897

PubMed ID

  • 19340143

Pubmed Central ID

  • 19340143

Electronic International Standard Serial Number (EISSN)

  • 1539-4522

International Standard Serial Number (ISSN)

  • 1559-128X

Digital Object Identifier (DOI)

  • 10.1364/ao.48.001891

Language

  • eng