Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks.

Publication ,  Journal Article
Silver, MT; Sterns, LD; Piccini, JP; Joung, B; Ching, C-K; Pickett, RA; Rabinovich, R; Liu, S; Peterson, BJ; Lexcen, DR; Shock-Less Investigators,
Published in: Heart Rhythm
March 2015

BACKGROUND: Implantable cardioverter-defibrillator (ICD) shocks are associated with increased anxiety, health care utilization, and potentially mortality. OBJECTIVE: The purpose of the Shock-Less Study was to determine if providing feedback reports to physicians on their adherence to evidence-based shock reduction programming could improve their programming behavior and reduce shocks. METHODS: Shock-Less enrolled primary prevention (PP) and secondary prevention (SP) ICD patients between 2009 and 2012 at 118 study centers worldwide and followed patients longitudinally after their ICD implant. Center-specific therapy programming reports (TPRs) were delivered to each center 9 to 12 months after their first enrollment. The reports detailed adherence to evidence-based programming targets: number of intervals to detect ventricular fibrillation (VF NID), longest treatment interval (LTI), supraventricular tachycardia (SVT) discriminators (Wavelet, PR Logic), SVT limit, Lead Integrity Alert (LIA), and antitachycardia pacing (ATP). Clinicians programmed ICDs at their discretion. The primary outcome measure was the change in utilization of evidence-based shock reduction programming before (phase I, n = 2694 patients) and after initiation of the TPR (phase II, n = 1438 patients). RESULTS: Patients implanted after feedback reports (phase II) were up to 20% more likely to have their ICDs programmed in line with evidence-based shock reduction programming (eg, VF NID in PP patients 30/40 in 33.5% vs 18.6%, P < .0001). Patients implanted in phase II had a lower risk of all-cause shock (adjusted hazard ratio 0.72, 95% confidence interval 0.58-0.90, P = .003). CONCLUSION: Providing programming feedback reports improves adherence to evidence-based shock reduction programming and is associated with lower risk of ICD shocks.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

March 2015

Volume

12

Issue

3

Start / End Page

545 / 553

Location

United States

Related Subject Headings

  • Ventricular Fibrillation
  • Treatment Outcome
  • Tachycardia, Supraventricular
  • Syncope
  • Secondary Prevention
  • Primary Prevention
  • Middle Aged
  • Male
  • Incidence
  • Humans
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Silver, M. T., Sterns, L. D., Piccini, J. P., Joung, B., Ching, C.-K., Pickett, R. A., … Shock-Less Investigators, . (2015). Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks. Heart Rhythm, 12(3), 545–553. https://doi.org/10.1016/j.hrthm.2014.11.002
Silver, Marc T., Laurence D. Sterns, Jonathan P. Piccini, Boyoung Joung, Chi-Keong Ching, Robert A. Pickett, Rafael Rabinovich, et al. “Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks.Heart Rhythm 12, no. 3 (March 2015): 545–53. https://doi.org/10.1016/j.hrthm.2014.11.002.
Silver MT, Sterns LD, Piccini JP, Joung B, Ching C-K, Pickett RA, et al. Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks. Heart Rhythm. 2015 Mar;12(3):545–53.
Silver, Marc T., et al. “Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks.Heart Rhythm, vol. 12, no. 3, Mar. 2015, pp. 545–53. Pubmed, doi:10.1016/j.hrthm.2014.11.002.
Silver MT, Sterns LD, Piccini JP, Joung B, Ching C-K, Pickett RA, Rabinovich R, Liu S, Peterson BJ, Lexcen DR, Shock-Less Investigators. Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks. Heart Rhythm. 2015 Mar;12(3):545–553.
Journal cover image

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

March 2015

Volume

12

Issue

3

Start / End Page

545 / 553

Location

United States

Related Subject Headings

  • Ventricular Fibrillation
  • Treatment Outcome
  • Tachycardia, Supraventricular
  • Syncope
  • Secondary Prevention
  • Primary Prevention
  • Middle Aged
  • Male
  • Incidence
  • Humans