Skip to main content
Journal cover image

Assessment of Device-Related Thrombus and Associated Clinical Outcomes With the WATCHMAN Left Atrial Appendage Closure Device for Embolic Protection in Patients With Atrial Fibrillation (from the PROTECT-AF Trial).

Publication ,  Journal Article
Main, ML; Fan, D; Reddy, VY; Holmes, DR; Gordon, NT; Coggins, TR; House, JA; Liao, L; Rabineau, D; Latus, GG; Huber, KC; Sievert, H ...
Published in: Am J Cardiol
April 1, 2016

Left atrial appendage closure with the WATCHMAN device is an alternative to anticoagulation for stroke prevention in selected patients with atrial fibrillation (AF). LA device-related thrombus (DRT) is poorly defined and understood. We aimed to (1) develop consensus echocardiographic diagnostic criteria for DRT; (2) estimate the incidence of DRT; and (3) determine clinical event rates in patients with DRT. In phase 1 (training), a training manual was developed and reviewed by 3 echocardiographers with left atrial appendage closure device experience. All available transesophageal (TEE) studies in the WATCHMAN left atrial appendage system for embolic protection in patients with atrial fibrillation (PROTECT-AF) trial patients with suspected DRT were reviewed in 2 subsequent phases. In phase 2 (primary blind read), each reviewer independently scored each study for DRT, and final echo criteria were developed. Unanimously scored studies were considered adjudicated, whereas all others were reevaluated by all reviewers in phase 3 (group adjudication read). DRT was suspected in 35 of 485 patients by the site investigator, the echocardiography core laboratory, or both; 93 of the individual TEE studies were available for review. In phase 2, 3 readers agreed on 67 (72%) of time points. Based on phases 1 and 2, 5 DRT criteria were developed. In phase 3, studies without agreement in phase 2 were adjudicated using these criteria. Overall, at least 1 TEE was DRT positive in 27 (5.7%) PROTECT-AF patients. Stroke, peripheral embolism, or cardiac/unexplained death occurred in subjects with DRT at a rate of 3.4 per 100 patient-years follow-up. In conclusion, DRT were identified on at least 1 TEE in 27 PROTECT-AF patients, indicating a DRT incidence of 5.7%. Primary efficacy events in patients with DRT occurred at a rate of 3.4 per 100 patient-years follow-up, intermediate in frequency between event rates previously reported for the overall device and warfarin arms in PROTECT-AF.

Duke Scholars

Published In

Am J Cardiol

DOI

EISSN

1879-1913

Publication Date

April 1, 2016

Volume

117

Issue

7

Start / End Page

1127 / 1134

Location

United States

Related Subject Headings

  • Warfarin
  • Treatment Outcome
  • Stroke
  • Middle Aged
  • Male
  • Incidence
  • Humans
  • Follow-Up Studies
  • Female
  • Embolic Protection Devices
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Main, M. L., Fan, D., Reddy, V. Y., Holmes, D. R., Gordon, N. T., Coggins, T. R., … Douglas, P. S. (2016). Assessment of Device-Related Thrombus and Associated Clinical Outcomes With the WATCHMAN Left Atrial Appendage Closure Device for Embolic Protection in Patients With Atrial Fibrillation (from the PROTECT-AF Trial). Am J Cardiol, 117(7), 1127–1134. https://doi.org/10.1016/j.amjcard.2016.01.039
Main, Michael L., Dali Fan, Vivek Y. Reddy, David R. Holmes, Nicole T. Gordon, Tina R. Coggins, John A. House, et al. “Assessment of Device-Related Thrombus and Associated Clinical Outcomes With the WATCHMAN Left Atrial Appendage Closure Device for Embolic Protection in Patients With Atrial Fibrillation (from the PROTECT-AF Trial).Am J Cardiol 117, no. 7 (April 1, 2016): 1127–34. https://doi.org/10.1016/j.amjcard.2016.01.039.
Main ML, Fan D, Reddy VY, Holmes DR, Gordon NT, Coggins TR, House JA, Liao L, Rabineau D, Latus GG, Huber KC, Sievert H, Wright RF, Doshi SK, Douglas PS. Assessment of Device-Related Thrombus and Associated Clinical Outcomes With the WATCHMAN Left Atrial Appendage Closure Device for Embolic Protection in Patients With Atrial Fibrillation (from the PROTECT-AF Trial). Am J Cardiol. 2016 Apr 1;117(7):1127–1134.
Journal cover image

Published In

Am J Cardiol

DOI

EISSN

1879-1913

Publication Date

April 1, 2016

Volume

117

Issue

7

Start / End Page

1127 / 1134

Location

United States

Related Subject Headings

  • Warfarin
  • Treatment Outcome
  • Stroke
  • Middle Aged
  • Male
  • Incidence
  • Humans
  • Follow-Up Studies
  • Female
  • Embolic Protection Devices