Skip to main content

Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant.

Publication ,  Journal Article
Anderson, DD; Tochigi, Y; Rudert, MJ; Vaseenon, T; Brown, TD; Amendola, A
Published in: J Bone Joint Surg Am
June 2010

BACKGROUND: Talar osteochondral defects can lead to joint degeneration. Focal resurfacing with a metallic implant has shown promise in other joints. We studied the effect of implantation accuracy on ankle contact mechanics after focal resurfacing of a defect in the talar dome. METHODS: Static loading of seven cadaver ankles was performed before and after creation of a 15-mm-diameter osteochondral defect on the talar dome, and joint contact stresses were measured. The defect was then resurfaced with a metallic implant, with use of a custom implant-bone interface fixture that allowed fine control (in 0.25-mm steps) of implantation height. Stress measurements were repeated at heights of -0.5 to +0.5 mm relative to an as-implanted reference. Finite element analysis was used to determine the effect of implant height, post axis rotation, and valgus/varus tilt over a motion duty cycle. RESULTS: With the untreated defect, there was a 20% reduction in contact area and a 40% increase in peak contact stress, as well as a shift in the location of the most highly loaded region, as compared with the values in the intact condition. Resurfacing led to recovery of 90% of the contact area that had been measured in the intact specimen, but the peak contact stresses remained elevated. With the implant 0.25 mm proud, peak contact stress was 220% of that in the intact specimen. The results of the finite element analyses agreed closely with those of the experiments and additionally showed substantial variations in defect influences on contact stresses across the motion arc. Talar internal/external rotations also differed for the unfilled defect. Focal implant resurfacing substantially restored kinematics but did not restore the stresses to the levels in the intact specimens. CONCLUSIONS: Focal resurfacing with a metallic implant appears to have the potential to restore normal joint mechanics in ankles with a large talar osteochondral defect. However, contact stresses were found to be highly sensitive to implant positioning.

Duke Scholars

Published In

J Bone Joint Surg Am

DOI

EISSN

1535-1386

Publication Date

June 2010

Volume

92

Issue

6

Start / End Page

1490 / 1500

Location

United States

Related Subject Headings

  • Stress, Mechanical
  • Orthopedics
  • Male
  • Joint Prosthesis
  • Humans
  • Female
  • Cadaver
  • Biomechanical Phenomena
  • Arthroplasty, Replacement
  • Ankle Joint
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Anderson, D. D., Tochigi, Y., Rudert, M. J., Vaseenon, T., Brown, T. D., & Amendola, A. (2010). Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant. J Bone Joint Surg Am, 92(6), 1490–1500. https://doi.org/10.2106/JBJS.I.00431
Anderson, Donald D., Yuki Tochigi, M James Rudert, Tanawat Vaseenon, Thomas D. Brown, and Annunziato Amendola. “Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant.J Bone Joint Surg Am 92, no. 6 (June 2010): 1490–1500. https://doi.org/10.2106/JBJS.I.00431.
Anderson DD, Tochigi Y, Rudert MJ, Vaseenon T, Brown TD, Amendola A. Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant. J Bone Joint Surg Am. 2010 Jun;92(6):1490–500.
Anderson, Donald D., et al. “Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant.J Bone Joint Surg Am, vol. 92, no. 6, June 2010, pp. 1490–500. Pubmed, doi:10.2106/JBJS.I.00431.
Anderson DD, Tochigi Y, Rudert MJ, Vaseenon T, Brown TD, Amendola A. Effect of implantation accuracy on ankle contact mechanics with a metallic focal resurfacing implant. J Bone Joint Surg Am. 2010 Jun;92(6):1490–1500.

Published In

J Bone Joint Surg Am

DOI

EISSN

1535-1386

Publication Date

June 2010

Volume

92

Issue

6

Start / End Page

1490 / 1500

Location

United States

Related Subject Headings

  • Stress, Mechanical
  • Orthopedics
  • Male
  • Joint Prosthesis
  • Humans
  • Female
  • Cadaver
  • Biomechanical Phenomena
  • Arthroplasty, Replacement
  • Ankle Joint