Skip to main content

A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging

Publication ,  Conference
Sahbaee, P; Abadi, E; Sanders, J; Becchetti, M; Zhang, Y; Agasthya, G; Segars, P; Samei, E
Published in: Progress in Biomedical Optics and Imaging - Proceedings of SPIE
January 1, 2016

The purpose of this study was to substantiate the interdependency of image quality, radiation dose, and contrast material dose in CT towards the patient-specific optimization of the imaging protocols. The study deployed two phantom platforms. First, a variable sized phantom containing an iodinated insert was imaged on a representative CT scanner at multiple CTDI values. The contrast and noise were measured from the reconstructed images for each phantom diameter. Linearly related to iodine-concentration, contrast to noise ratio (CNR), was calculated for different iodine-concentration levels. Second, the analysis was extended to a recently developed suit of 58 virtual human models (5D-XCAT) with added contrast dynamics. Emulating a contrast-enhanced abdominal image procedure and targeting a peak-enhancement in aorta, each XCAT phantom was "imaged" using a CT simulation platform. 3D surfaces for each patient/size established the relationship between iodine-concentration, dose, and CNR. The Sensitivity of Ratio (SR), defined as ratio of change in iodine-concentration versus dose to yield a constant change in CNR was calculated and compared at high and low radiation dose for both phantom platforms. The results show that sensitivity of CNR to iodine concentration is larger at high radiation dose (up to 73%). The SR results were highly affected by radiation dose metric; CTDI or organ dose. Furthermore, results showed that the presence of contrast material could have a profound impact on optimization results (up to 45%).

Duke Scholars

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

Publication Date

January 1, 2016

Volume

9783
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Sahbaee, P., Abadi, E., Sanders, J., Becchetti, M., Zhang, Y., Agasthya, G., … Samei, E. (2016). A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging. In Progress in Biomedical Optics and Imaging - Proceedings of SPIE (Vol. 9783). https://doi.org/10.1117/12.2216516
Sahbaee, P., E. Abadi, J. Sanders, M. Becchetti, Y. Zhang, G. Agasthya, P. Segars, and E. Samei. “A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging.” In Progress in Biomedical Optics and Imaging - Proceedings of SPIE, Vol. 9783, 2016. https://doi.org/10.1117/12.2216516.
Sahbaee P, Abadi E, Sanders J, Becchetti M, Zhang Y, Agasthya G, et al. A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging. In: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2016.
Sahbaee, P., et al. “A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging.” Progress in Biomedical Optics and Imaging - Proceedings of SPIE, vol. 9783, 2016. Scopus, doi:10.1117/12.2216516.
Sahbaee P, Abadi E, Sanders J, Becchetti M, Zhang Y, Agasthya G, Segars P, Samei E. A technique for multi-dimensional optimization of radiation dose, contrast dose, and image quality in CT imaging. Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2016.

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

Publication Date

January 1, 2016

Volume

9783