Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.

Publication ,  Journal Article
Lam, DSC; Leung, DYL; Chiu, TYH; Fan, DSP; Cheung, EYY; Wong, T-Y; Lai, JSM; Tham, CCY
Published in: Invest Ophthalmol Vis Sci
September 2004

PURPOSE: To evaluate whether the pressure phosphene tonometer (PPT) is suitable for self-tonometry in patients with glaucoma or ocular hypertension. METHODS: This was a prospective comparative study of 102 eyes of 102 patients with chronic glaucoma or ocular hypertension. Intraocular pressure (IOP) measurements by the Goldmann tonometer (GT) were compared with self-measured readings with the PPT. Patients evaluated the ease of home use of the PPT. The last 15 patients were asked to stop their glaucoma medications, and the ability of the PPT to detect an elevated IOP during self-tonometry was studied. RESULTS: The mean +/- SD difference between PPT and GT readings was -0.24 +/- 1.57 mm Hg. Comparing the PPT with the GT, 86% of the readings were within +/- 2.0 mm Hg, and 91% were within +/- 3.0 mm Hg. Spearman's correlation coefficient was 0.91. A Bland-Altman plot showed that the 95% limits of agreement between the two methods lay between 2.90 and -3.38 mm Hg. Within-subject coefficients of variation for the GT and the PPT were 4.4% and 7.3%, respectively. In detecting an elevated IOP of more than 21 mm Hg, the sensitivity and the specificity of the PPT were 92.3% and 98.6% respectively. The mean satisfaction score for home use of the PPT was 87.4 +/- 16.3 (maximum 100). CONCLUSIONS: With proper training and technique, self-tonometry with the PPT appears to be accurate up to at least 25 mm Hg and is reproducible. The PPT was sensitive and specific in detecting an elevated IOP of more than 21 mm Hg. As patients were expected to seek ophthalmic care before the self-measured IOP reaches 25 mm Hg, the PPT may have a value for self-monitoring. Patients rated the PPT as satisfactory for home use. Because the PPT is portable and relatively inexpensive and requires no topical anesthesia or direct contact with the eyeball, it may have potential as an instrument for home self-tonometry.

Duke Scholars

Published In

Invest Ophthalmol Vis Sci

DOI

ISSN

0146-0404

Publication Date

September 2004

Volume

45

Issue

9

Start / End Page

3131 / 3136

Location

United States

Related Subject Headings

  • Tonometry, Ocular
  • Sensitivity and Specificity
  • Self Care
  • Reproducibility of Results
  • Prospective Studies
  • Phosphenes
  • Patient Satisfaction
  • Ophthalmology & Optometry
  • Ocular Hypertension
  • Middle Aged
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Lam, D. S. C., Leung, D. Y. L., Chiu, T. Y. H., Fan, D. S. P., Cheung, E. Y. Y., Wong, T.-Y., … Tham, C. C. Y. (2004). Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci, 45(9), 3131–3136. https://doi.org/10.1167/iovs.04-0115
Lam, Dennis S. C., Dexter Y. L. Leung, Thomas Y. H. Chiu, Dorothy S. P. Fan, Eva Y. Y. Cheung, Tien-Yin Wong, Jimmy S. M. Lai, and Clement C. Y. Tham. “Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.Invest Ophthalmol Vis Sci 45, no. 9 (September 2004): 3131–36. https://doi.org/10.1167/iovs.04-0115.
Lam DSC, Leung DYL, Chiu TYH, Fan DSP, Cheung EYY, Wong T-Y, et al. Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci. 2004 Sep;45(9):3131–6.
Lam, Dennis S. C., et al. “Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients.Invest Ophthalmol Vis Sci, vol. 45, no. 9, Sept. 2004, pp. 3131–36. Pubmed, doi:10.1167/iovs.04-0115.
Lam DSC, Leung DYL, Chiu TYH, Fan DSP, Cheung EYY, Wong T-Y, Lai JSM, Tham CCY. Pressure phosphene self-tonometry: a comparison with goldmann tonometry in glaucoma patients. Invest Ophthalmol Vis Sci. 2004 Sep;45(9):3131–3136.

Published In

Invest Ophthalmol Vis Sci

DOI

ISSN

0146-0404

Publication Date

September 2004

Volume

45

Issue

9

Start / End Page

3131 / 3136

Location

United States

Related Subject Headings

  • Tonometry, Ocular
  • Sensitivity and Specificity
  • Self Care
  • Reproducibility of Results
  • Prospective Studies
  • Phosphenes
  • Patient Satisfaction
  • Ophthalmology & Optometry
  • Ocular Hypertension
  • Middle Aged