Post-market surveillance to detect adverse events associated with Melody® valve implantation.

Publication ,  Journal Article
Hill, KD; Goldstein, BH; Angtuaco, MJ; Chu, PY; Fleming, GA
Published in: Cardiol Young
August 2017

OBJECTIVE: The aim of this study was to describe previously unrecognised or under-recognised adverse events associated with Melody® valve implantation. BACKGROUND: In rare diseases and conditions, it is typically not feasible to conduct large-scale safety trials before drug or device approval. Therefore, post-market surveillance mechanisms are necessary to detect rare but potentially serious adverse events. METHODS: We reviewed the United States Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database and conducted a structured literature review to evaluate adverse events associated with on- and off-label Melody® valve implantation. Adverse events were compared with those described in the prospective Investigational Device Exemption and Post-Market Approval Melody® transcatheter pulmonary valve trials. RESULTS: We identified 631 adverse events associated with "on-label" Melody® valve implants and 84 adverse events associated with "off-label" implants. The most frequent "on-label" adverse events were similar to those described in the prospective trials including stent fracture (n=210) and endocarditis (n=104). Previously unrecognised or under-recognised adverse events included stent fragment embolisation (n=5), device erosion (n=4), immediate post-implant severe valvar insufficiency (n=2), and late coronary compression (n=2 cases at 5 days and 3 months after implantation). Under-recognised adverse events associated with off-label implantation included early valve failure due to insufficiency when implanted in the tricuspid position (n=7) and embolisation with percutaneous implantation in the mitral position (n=5). CONCLUSION: Post-market passive surveillance does not demonstrate a high frequency of previously unrecognised serious adverse events with "on-label" Melody® valve implantation. Further study is needed to evaluate safety of "off-label" uses.

Duke Scholars

Published In

Cardiol Young

DOI

EISSN

1467-1107

Publication Date

August 2017

Volume

27

Issue

6

Start / End Page

1090 / 1097

Location

England

Related Subject Headings

  • United States Food and Drug Administration
  • United States
  • Pulmonary Valve
  • Prosthesis Failure
  • Prosthesis Design
  • Product Surveillance, Postmarketing
  • Humans
  • Heart Valve Prosthesis
  • Heart Valve Diseases
  • Cardiovascular System & Hematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Hill, K. D., Goldstein, B. H., Angtuaco, M. J., Chu, P. Y., & Fleming, G. A. (2017). Post-market surveillance to detect adverse events associated with Melody® valve implantation. Cardiol Young, 27(6), 1090–1097. https://doi.org/10.1017/S1047951116002092
Hill, Kevin D., Bryan H. Goldstein, Michael J. Angtuaco, Patricia Y. Chu, and Gregory A. Fleming. “Post-market surveillance to detect adverse events associated with Melody® valve implantation.Cardiol Young 27, no. 6 (August 2017): 1090–97. https://doi.org/10.1017/S1047951116002092.
Hill KD, Goldstein BH, Angtuaco MJ, Chu PY, Fleming GA. Post-market surveillance to detect adverse events associated with Melody® valve implantation. Cardiol Young. 2017 Aug;27(6):1090–7.
Hill, Kevin D., et al. “Post-market surveillance to detect adverse events associated with Melody® valve implantation.Cardiol Young, vol. 27, no. 6, Aug. 2017, pp. 1090–97. Pubmed, doi:10.1017/S1047951116002092.
Hill KD, Goldstein BH, Angtuaco MJ, Chu PY, Fleming GA. Post-market surveillance to detect adverse events associated with Melody® valve implantation. Cardiol Young. 2017 Aug;27(6):1090–1097.
Journal cover image

Published In

Cardiol Young

DOI

EISSN

1467-1107

Publication Date

August 2017

Volume

27

Issue

6

Start / End Page

1090 / 1097

Location

England

Related Subject Headings

  • United States Food and Drug Administration
  • United States
  • Pulmonary Valve
  • Prosthesis Failure
  • Prosthesis Design
  • Product Surveillance, Postmarketing
  • Humans
  • Heart Valve Prosthesis
  • Heart Valve Diseases
  • Cardiovascular System & Hematology