SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods.

Publication ,  Journal Article
Yang, Y; Cai, J; Meltsner, S; Chang, Z; Craciunescu, O
Published in: Med Phys
June 2016

PURPOSE: The Varian tandem and ring applicators are used to deliver HDR Ir-192 brachytherapy for cervical cancer. The source path within the ring is hard to predict due to the larger interior ring lumen. Some studies showed the source could be several millimeters different from planned positions, while other studies demonstrated minimal dosimetric impact. A global shift can be applied to limit the effect of positioning offsets. The purpose of this study was to assess the necessities of implementing a global source shift using Monte Carlo (MC) simulations. METHODS: The MCNP5 radiation transport code was used for all MC simulations. To accommodate TG-186 guidelines and eliminate inter-source attenuation, a BrachyVision plan with 10 dwell positions (0.5cm step sizes) was simulated as the summation of 10 individual sources with equal dwell times for simplification. To simplify the study, the tandem was also excluded from the MC model. Global shifts of ±0.1, ±0.3, ±0.5 cm were then simulated as distal and proximal from the reference positions. Dose was scored in water for all MC simulations and was normalized to 100% at the normalization point 0.5 cm from the cap in the ring plane. For dose comparison, Point A was 2 cm caudal from the buildup cap and 2 cm lateral on either side of the ring axis. With seventy simulations, 108 photon histories gave a statistical uncertainties (k=1) <2% for (0.1 cm)3 voxels. RESULTS: Compared to no global shift, average Point A doses were 0.0%, 0.4%, and 2.2% higher for distal global shifts, and 0.4%, 2.8%, and 5.1% higher for proximal global shifts, respectively. The MC Point A doses differed by < 1% when compared to BrachyVision. CONCLUSION: Dose variations were not substantial for ±0.3 cm global shifts, which is common in clinical practice.

Duke Scholars

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

June 2016

Volume

43

Issue

6

Start / End Page

3625 / 3626

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Yang, Y., Cai, J., Meltsner, S., Chang, Z., & Craciunescu, O. (2016). SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods. Med Phys, 43(6), 3625–3626. https://doi.org/10.1118/1.4956886
Yang, Y., J. Cai, S. Meltsner, Z. Chang, and O. Craciunescu. “SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods.Med Phys 43, no. 6 (June 2016): 3625–26. https://doi.org/10.1118/1.4956886.
Yang, Y., et al. “SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods.Med Phys, vol. 43, no. 6, June 2016, pp. 3625–26. Pubmed, doi:10.1118/1.4956886.

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

June 2016

Volume

43

Issue

6

Start / End Page

3625 / 3626

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences