Estimating detectability index in vivo: development and validation of an automated methodology.
This study's purpose was to develop and validate a method to estimate patient-specific detectability indices directly from patients' CT images (i.e., in vivo). The method extracts noise power spectrum (NPS) and modulation transfer function (MTF) resolution properties from each patient's CT series based on previously validated techniques. These are combined with a reference task function (10-mm disk lesion with [Formula: see text] HU contrast) to estimate detectability indices for a nonprewhitening matched filter observer model. This method was applied to CT data from a previous study in which diagnostic performance of 16 readers was measured for the task of detecting subtle, hypoattenuating liver lesions ([Formula: see text]), using a two-alternative-forced-choice (2AFC) method, over six dose levels and two reconstruction algorithms. In vivo detectability indices were estimated and compared to the human readers' binary 2AFC outcomes using a generalized linear mixed-effects statistical model. The results of this modeling showed that the in vivo detectability indices were strongly related to 2AFC outcomes ([Formula: see text]). Linear comparison between human-detection accuracy and model-predicted detection accuracy (for like conditions) resulted in Pearson and Spearman correlation coefficients exceeding 0.84. These results suggest the potential utility of using in vivo estimates of a detectability index for an automated image quality tracking system that could be implemented clinically.
Duke Scholars
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- 4003 Biomedical engineering
- 3202 Clinical sciences
Citation
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- 4003 Biomedical engineering
- 3202 Clinical sciences