How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study.

Publication ,  Journal Article
Euler, A; Solomon, J; Mazurowski, MA; Samei, E; Nelson, RC
Published in: Eur Radiol
April 2019

OBJECTIVES: To assess the impact of scan- and patient-related factors on the error and the minimum detectable difference in iodine concentration among different generations of single-source (SS) fast kV-switching and dual-source (DS) dual-energy CT (DECT). METHODS: Lesions having eight different iodine concentrations (0.2-4 mgI/mL) were emulated in a 3D-printed phantom of medium and large size. Each combination of concentration and size was scanned in dual-energy mode on four different SS and DS DECTs. Radiation doses were 7 and 10 mGy (medium size) and 10, 13, and 16 mGy (large size). Iodine maps were reconstructed with filtered back projection (FBP) and vendor-specific iterative reconstruction algorithms (IRs). Absolute error of iodine quantification (E) was measured. Multivariate regression models determined the influence of CT scanner, iodine concentration, phantom size, radiation dose, and reconstruction algorithm on E. The minimum detectable difference in iodine concentration (ICmin) under the same imaging conditions (intra-conditional) and among different imaging conditions (inter-conditional) was calculated. RESULTS: The error was significantly lower in current than in previous DECT generations (p < 0.001). For all CT scanner conditions, the error was significantly higher with increasing phantom size and decreasing radiation dose (p < 0.001). Iodine concentration only significantly affected the error for SS DECT (p < 0.001). ICmin depended on patient- and scan-related factors and ranged from 0.4 to 1.5 mgI/mL. CONCLUSIONS: Patient- and scan-related factors have a significant impact on the error and minimum detectable difference in iodine concentration within and among SS fast kV-switching and DS DECT. KEY POINTS: • Patient- and scan-related factors have a significant impact on the error and minimum detectable difference in dual-energy CT-based iodine quantification. • Third-generation DECTs outperformed second-generation scanners for both single-source and dual-source dual-energy CT. • The minimum intra- and inter-conditional detectable difference in iodine concentration ranged from 0.4 to 1.5 mg iodine/mL.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Eur Radiol

DOI

EISSN

1432-1084

Publication Date

April 2019

Volume

29

Issue

4

Start / End Page

2069 / 2078

Location

Germany

Related Subject Headings

  • Reproducibility of Results
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Multidetector Computed Tomography
  • Iodine
  • Humans
  • Algorithms
  • 3202 Clinical sciences
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Euler, A., Solomon, J., Mazurowski, M. A., Samei, E., & Nelson, R. C. (2019). How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study. Eur Radiol, 29(4), 2069–2078. https://doi.org/10.1007/s00330-018-5736-0
Euler, André, Justin Solomon, Maciej A. Mazurowski, Ehsan Samei, and Rendon C. Nelson. “How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study.Eur Radiol 29, no. 4 (April 2019): 2069–78. https://doi.org/10.1007/s00330-018-5736-0.
Journal cover image

Published In

Eur Radiol

DOI

EISSN

1432-1084

Publication Date

April 2019

Volume

29

Issue

4

Start / End Page

2069 / 2078

Location

Germany

Related Subject Headings

  • Reproducibility of Results
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Multidetector Computed Tomography
  • Iodine
  • Humans
  • Algorithms
  • 3202 Clinical sciences
  • 1103 Clinical Sciences