Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay.

Publication ,  Journal Article
Fitzgerald, RL; Hollander, JE; Peacock, WF; Limkakeng, AT; Breitenbeck, N; Blechschmidt, K; Laimighofer, M; deFilippi, C
Published in: Clin Chim Acta
August 2019

BACKGROUND: We report the analytical performance of the Elecsys® Troponin T Gen 5 STAT (TnT Gen 5 STAT; Roche Diagnostics) assay. METHODS: Measuring limits/ranges were determined in lithium-heparin plasma samples per Clinical and Laboratory Standards Institute (CLSI) EP17-A2. Precision was evaluated per CLSI EP05-A2 using lithium-heparin plasma/quality control samples on cobas e 411/cobas e 601 analyzers; two duplicated runs per day for 21 days (n = 84). Cross-reactivity with other troponin forms and interference from endogenous substances/drugs was tested; recovery criterion for no cross-reactivity was within ±10%. RESULTS: Coefficients of variation (CV) for repeatability/intermediate precision were 0.7-5.6%/1.4-10.3% (cobas e 411; mean cardiac troponin T [cTnT]: 7.3-9341 ng/L) and 0.7-3.0%/1.5-6.4% (cobas e 601; mean cTnT: 7.4-9455 ng/L). There was no cross-reactivity with skeletal muscle troponin T (≤ 10,000 ng/L), skeletal muscle troponin I (≤ 100,000 ng/L), cardiac troponin I (≤ 10,000 ng/L), or human troponin C (≤ 80,000 ng/L). No interference was observed with biotin (≤ 20 ng/mL) or 34 drugs. CONCLUSION: The TnT Gen 5 STAT assay demonstrated a CV of <10% at the 99th percentile upper reference limit, meeting precision requirements (Fourth Universal Definition of Myocardial Infarction) for high-sensitivity troponin assays.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Clin Chim Acta

DOI

EISSN

1873-3492

Publication Date

August 2019

Volume

495

Start / End Page

522 / 528

Location

Netherlands

Related Subject Headings

  • Troponin T
  • Reproducibility of Results
  • Luminescent Measurements
  • Limit of Detection
  • Immunoassay
  • Humans
  • General Clinical Medicine
  • Blood Chemical Analysis
  • 3202 Clinical sciences
  • 3101 Biochemistry and cell biology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Fitzgerald, R. L., Hollander, J. E., Peacock, W. F., Limkakeng, A. T., Breitenbeck, N., Blechschmidt, K., … deFilippi, C. (2019). Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay. Clin Chim Acta, 495, 522–528. https://doi.org/10.1016/j.cca.2019.05.026
Fitzgerald, Robert L., Judd E. Hollander, W Frank Peacock, Alexander T. Limkakeng, Nancy Breitenbeck, Kareen Blechschmidt, Michael Laimighofer, and Christopher deFilippi. “Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay.Clin Chim Acta 495 (August 2019): 522–28. https://doi.org/10.1016/j.cca.2019.05.026.
Fitzgerald RL, Hollander JE, Peacock WF, Limkakeng AT, Breitenbeck N, Blechschmidt K, et al. Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay. Clin Chim Acta. 2019 Aug;495:522–8.
Fitzgerald, Robert L., et al. “Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay.Clin Chim Acta, vol. 495, Aug. 2019, pp. 522–28. Pubmed, doi:10.1016/j.cca.2019.05.026.
Fitzgerald RL, Hollander JE, Peacock WF, Limkakeng AT, Breitenbeck N, Blechschmidt K, Laimighofer M, deFilippi C. Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay. Clin Chim Acta. 2019 Aug;495:522–528.
Journal cover image

Published In

Clin Chim Acta

DOI

EISSN

1873-3492

Publication Date

August 2019

Volume

495

Start / End Page

522 / 528

Location

Netherlands

Related Subject Headings

  • Troponin T
  • Reproducibility of Results
  • Luminescent Measurements
  • Limit of Detection
  • Immunoassay
  • Humans
  • General Clinical Medicine
  • Blood Chemical Analysis
  • 3202 Clinical sciences
  • 3101 Biochemistry and cell biology