Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings.

Publication ,  Journal Article
Meyer, M; Ronald, J; Vernuccio, F; Nelson, RC; Ramirez-Giraldo, JC; Solomon, J; Patel, BN; Samei, E; Marin, D
Published in: Radiology
December 2019

Background Results of recent phantom studies show that variation in CT acquisition parameters and reconstruction techniques may make radiomic features largely nonreproduceable and of limited use for prognostic clinical studies. Purpose To investigate the effect of CT radiation dose and reconstruction settings on the reproducibility of radiomic features, as well as to identify correction factors for mitigating these sources of variability. Materials and Methods This was a secondary analysis of a prospective study of metastatic liver lesions in patients who underwent staging with single-energy dual-source contrast material-enhanced staging CT between September 2011 and April 2012. Technique parameters were altered, resulting in 28 CT data sets per patient that included different dose levels, section thicknesses, kernels, and reconstruction algorithm settings. By using a training data set (n = 76), reproducible intensity, shape, and texture radiomic features (reproducibility threshold, R2 ≥ 0.95) were selected and correction factors were calculated by using a linear model to convert each radiomic feature to its estimated value in a reference technique. By using a test data set (n = 75), the reproducibility of hierarchical clustering based on 106 radiomic features measured with different CT techniques was assessed. Results Data in 78 patients (mean age, 60 years ± 10; 33 women) with 151 liver lesions were included. The percentage of radiomic features deemed reproducible for any variation of the different technical parameters was 11% (12 of 106). Of all technical parameters, reconstructed section thickness had the largest impact on the reproducibility of radiomic features (12.3% [13 of 106]) if only one technical parameter was changed while all other technical parameters were kept constant. The results of the hierarchical cluster analysis showed improved clustering reproducibility when reproducible radiomic features with dedicated correction factors were used (ρ = 0.39-0.71 vs ρ = 0.14-0.47). Conclusion Most radiomic features are highly affected by CT acquisition and reconstruction settings, to the point of being nonreproducible. Selecting reproducible radiomic features along with study-specific correction factors offers improved clustering reproducibility. © RSNA, 2019 Online supplemental material is available for this article. See also the editorial by Sosna in this issue.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Radiology

DOI

EISSN

1527-1315

Publication Date

December 2019

Volume

293

Issue

3

Start / End Page

583 / 591

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Reproducibility of Results
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiation Dosage
  • Prospective Studies
  • Nuclear Medicine & Medical Imaging
  • Neoplasm Staging
  • Middle Aged
  • Male
  • Liver Neoplasms
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Meyer, M., Ronald, J., Vernuccio, F., Nelson, R. C., Ramirez-Giraldo, J. C., Solomon, J., … Marin, D. (2019). Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings. Radiology, 293(3), 583–591. https://doi.org/10.1148/radiol.2019190928
Meyer, Mathias, James Ronald, Federica Vernuccio, Rendon C. Nelson, Juan Carlos Ramirez-Giraldo, Justin Solomon, Bhavik N. Patel, Ehsan Samei, and Daniele Marin. “Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings.Radiology 293, no. 3 (December 2019): 583–91. https://doi.org/10.1148/radiol.2019190928.
Meyer M, Ronald J, Vernuccio F, Nelson RC, Ramirez-Giraldo JC, Solomon J, et al. Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings. Radiology. 2019 Dec;293(3):583–91.
Meyer, Mathias, et al. “Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings.Radiology, vol. 293, no. 3, Dec. 2019, pp. 583–91. Pubmed, doi:10.1148/radiol.2019190928.
Meyer M, Ronald J, Vernuccio F, Nelson RC, Ramirez-Giraldo JC, Solomon J, Patel BN, Samei E, Marin D. Reproducibility of CT Radiomic Features within the Same Patient: Influence of Radiation Dose and CT Reconstruction Settings. Radiology. 2019 Dec;293(3):583–591.

Published In

Radiology

DOI

EISSN

1527-1315

Publication Date

December 2019

Volume

293

Issue

3

Start / End Page

583 / 591

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Reproducibility of Results
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiation Dosage
  • Prospective Studies
  • Nuclear Medicine & Medical Imaging
  • Neoplasm Staging
  • Middle Aged
  • Male
  • Liver Neoplasms