Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study.

Publication ,  Journal Article
Garweg, C; Khelae, SK; Steinwender, C; Chan, JYS; Ritter, P; Johansen, JB; Sagi, V; Epstein, LM; Piccini, JP; Pascual, M; Mont, L; Willems, R ...
Published in: Heart Rhythm
December 2020

BACKGROUND: The MARVEL (Micra Atrial TRacking Using a Ventricular AccELerometer) 2 study assessed the efficacy of atrioventricular (AV) synchronous pacing with a Micra leadless pacemaker. Average atrioventricular synchrony (AVS) was 89.2%. Previously, low amplitude of the Micra-sensed atrial signal (A4) was observed to be a factor of low AVS. OBJECTIVE: The purpose of this study was to identify predictors of A4 amplitude and high AVS. METHODS: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS. RESULTS: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and e'/a' ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A <0.94 and SDSD <5 bpm. CONCLUSION: Clinical parameters and echocardiographic markers of atrial function are associated with A4 signal amplitude. High AVS can be predicted by E/A ratio <0.94 and low sinus rate variability at rest.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

December 2020

Volume

17

Issue

12

Start / End Page

2037 / 2045

Location

United States

Related Subject Headings

  • Young Adult
  • Prospective Studies
  • Pacemaker, Artificial
  • Middle Aged
  • Male
  • Humans
  • Heart Ventricles
  • Heart Atria
  • Follow-Up Studies
  • Female
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Garweg, C., Khelae, S. K., Steinwender, C., Chan, J. Y. S., Ritter, P., Johansen, J. B., … Chinitz, L. (2020). Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study. Heart Rhythm, 17(12), 2037–2045. https://doi.org/10.1016/j.hrthm.2020.07.024
Garweg, Christophe, Surinder Kaur Khelae, Clemens Steinwender, Joseph Yat Sun Chan, Philippe Ritter, Jens Brock Johansen, Venkata Sagi, et al. “Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study.Heart Rhythm 17, no. 12 (December 2020): 2037–45. https://doi.org/10.1016/j.hrthm.2020.07.024.
Garweg C, Khelae SK, Steinwender C, Chan JYS, Ritter P, Johansen JB, et al. Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study. Heart Rhythm. 2020 Dec;17(12):2037–45.
Garweg, Christophe, et al. “Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study.Heart Rhythm, vol. 17, no. 12, Dec. 2020, pp. 2037–45. Pubmed, doi:10.1016/j.hrthm.2020.07.024.
Garweg C, Khelae SK, Steinwender C, Chan JYS, Ritter P, Johansen JB, Sagi V, Epstein LM, Piccini JP, Pascual M, Mont L, Willems R, Sheldon T, Splett V, Stromberg K, Wood N, Chinitz L. Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study. Heart Rhythm. 2020 Dec;17(12):2037–2045.
Journal cover image

Published In

Heart Rhythm

DOI

EISSN

1556-3871

Publication Date

December 2020

Volume

17

Issue

12

Start / End Page

2037 / 2045

Location

United States

Related Subject Headings

  • Young Adult
  • Prospective Studies
  • Pacemaker, Artificial
  • Middle Aged
  • Male
  • Humans
  • Heart Ventricles
  • Heart Atria
  • Follow-Up Studies
  • Female