Evaluation of a quality control phantom for digital chest radiography

Publication ,  Journal Article
Mah, E; Samei, E; Peck, DJ
Published in: Journal of Applied Clinical Medical Physics
March 2001

: To examine the effectiveness and suitability of a quality control (QC) phantom for a routine QC program in digital radiography. : The chest phantom consists of copper and aluminum cutouts arranged to resemble the appearance of a chest. Performance of the digital radiography (DR) system is evaluated using high and low contrast resolution objects placed in the “heart,” “lung,” and “subdiaphragm” areas of the phantom. In addition, the signal levels from these areas were compared to similar areas from clinical chest radiographs. : The test objects included within the phantom were effective in assessing image quality except within the subdiaphragm area, where most of the low contrast disks were visible. Spatial resolution for the DR systems evaluated with the phantom ranged from 2.6 lp/mm to 4 lp/mm, falling within the middle of the line pair range provided. The signal levels of the heart and diaphragm regions relative to the lung region of the phantom were significantly higher than in clinical chest radiographs (0.67 versus 0.21 and 0.28 versus 0.10 for the heart and diaphragm regions, respectively). The heart‐to‐diaphragm signal level ratio, however, was comparable to those in clinical radiographs. : The findings suggest that the attenuation characteristics of the phantom are somewhat different from actual chests, but this did not appear to affect the post‐processing used by the imaging systems and usefulness for QC of these systems. The qualitative and quantitative measurements on the phantom for different systems were similar, suggesting that a single phantom can be used to evaluate system performance in a routine QC program for a wide range of digital radiography systems. This makes the implementation of a uniform QC program easier for institutions with a mixture of different digital radiography systems.PACS number(s): 87.57.–s, 87.62.+n

Duke Scholars

Published In

Journal of Applied Clinical Medical Physics

DOI

EISSN

1526-9914

ISSN

1526-9914

Publication Date

March 2001

Volume

2

Issue

2

Start / End Page

90 / 101

Publisher

Wiley

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 3208 Medical physiology
  • 1116 Medical Physiology
  • 1103 Clinical Sciences
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Mah, E., Samei, E., & Peck, D. J. (2001). Evaluation of a quality control phantom for digital chest radiography. Journal of Applied Clinical Medical Physics, 2(2), 90–101. https://doi.org/10.1120/jacmp.v2i2.2621
Mah, Eugene, Ehsan Samei, and Donald J. Peck. “Evaluation of a quality control phantom for digital chest radiography.” Journal of Applied Clinical Medical Physics 2, no. 2 (March 2001): 90–101. https://doi.org/10.1120/jacmp.v2i2.2621.
Mah E, Samei E, Peck DJ. Evaluation of a quality control phantom for digital chest radiography. Journal of Applied Clinical Medical Physics. 2001 Mar;2(2):90–101.
Mah, Eugene, et al. “Evaluation of a quality control phantom for digital chest radiography.” Journal of Applied Clinical Medical Physics, vol. 2, no. 2, Wiley, Mar. 2001, pp. 90–101. Crossref, doi:10.1120/jacmp.v2i2.2621.
Mah E, Samei E, Peck DJ. Evaluation of a quality control phantom for digital chest radiography. Journal of Applied Clinical Medical Physics. Wiley; 2001 Mar;2(2):90–101.

Published In

Journal of Applied Clinical Medical Physics

DOI

EISSN

1526-9914

ISSN

1526-9914

Publication Date

March 2001

Volume

2

Issue

2

Start / End Page

90 / 101

Publisher

Wiley

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 3208 Medical physiology
  • 1116 Medical Physiology
  • 1103 Clinical Sciences
  • 0299 Other Physical Sciences