Skip to main content
Journal cover image

Fully printed prothrombin time sensor for point-of-care testing.

Publication ,  Journal Article
Williams, NX; Carroll, B; Noyce, SG; Hobbie, HA; Joh, DY; Rogers, JG; Franklin, AD
Published in: Biosens Bioelectron
January 15, 2021

With an increasing number of patients relying on blood thinners to treat medical conditions, there is a rising need for rapid, low-cost, portable testing of blood coagulation time or prothrombin time (PT). Current methods for measuring PT require regular visits to outpatient clinics, which is cumbersome and time-consuming, decreasing patient quality of life. In this work, we developed a handheld point-of-care test (POCT) to measure PT using electrical transduction. Low-cost PT sensors were fully printed using an aerosol jet printer and conductive inks of Ag nanoparticles, Ag nanowires, and carbon nanotubes. Using benchtop control electronics to test this impedance-based biosensor, it was found that the capacitive nature of blood obscures the clotting response at frequencies below 10 kHz, leading to an optimized operating frequency of 15 kHz. When printed on polyimide, the PT sensor exhibited no variation in the measured clotting time, even when flexed to a 35 mm bend radius. In addition, consistent PT measurements for both chicken and human blood illustrate the versatility of these printed biosensors under disparate operating conditions, where chicken blood clots within 30 min and anticoagulated human blood clots within 20-100 s. Finally, a low-cost, handheld POCT was developed to measure PT for human blood, yielding 70% lower noise compared to measurement with a commercial potentiostat. This POCT with printed PT sensors has the potential to dramatically improve the quality of life for patients on blood thinners and, in the long term, could be incorporated into a fully flexible and wearable sensing platform.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Biosens Bioelectron

DOI

EISSN

1873-4235

Publication Date

January 15, 2021

Volume

172

Start / End Page

112770

Location

England

Related Subject Headings

  • Silver
  • Quality of Life
  • Prothrombin Time
  • Point-of-Care Testing
  • Point-of-Care Systems
  • Nanotubes, Carbon
  • Metal Nanoparticles
  • Humans
  • Biosensing Techniques
  • Bioinformatics
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Williams, N. X., Carroll, B., Noyce, S. G., Hobbie, H. A., Joh, D. Y., Rogers, J. G., & Franklin, A. D. (2021). Fully printed prothrombin time sensor for point-of-care testing. Biosens Bioelectron, 172, 112770. https://doi.org/10.1016/j.bios.2020.112770
Williams, Nicholas X., Brittani Carroll, Steven G. Noyce, Hansel Alex Hobbie, Daniel Y. Joh, Joseph G. Rogers, and Aaron D. Franklin. “Fully printed prothrombin time sensor for point-of-care testing.Biosens Bioelectron 172 (January 15, 2021): 112770. https://doi.org/10.1016/j.bios.2020.112770.
Williams NX, Carroll B, Noyce SG, Hobbie HA, Joh DY, Rogers JG, et al. Fully printed prothrombin time sensor for point-of-care testing. Biosens Bioelectron. 2021 Jan 15;172:112770.
Williams, Nicholas X., et al. “Fully printed prothrombin time sensor for point-of-care testing.Biosens Bioelectron, vol. 172, Jan. 2021, p. 112770. Pubmed, doi:10.1016/j.bios.2020.112770.
Williams NX, Carroll B, Noyce SG, Hobbie HA, Joh DY, Rogers JG, Franklin AD. Fully printed prothrombin time sensor for point-of-care testing. Biosens Bioelectron. 2021 Jan 15;172:112770.
Journal cover image

Published In

Biosens Bioelectron

DOI

EISSN

1873-4235

Publication Date

January 15, 2021

Volume

172

Start / End Page

112770

Location

England

Related Subject Headings

  • Silver
  • Quality of Life
  • Prothrombin Time
  • Point-of-Care Testing
  • Point-of-Care Systems
  • Nanotubes, Carbon
  • Metal Nanoparticles
  • Humans
  • Biosensing Techniques
  • Bioinformatics