Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®.

Publication ,  Journal Article
Ho, LM; Pendse, AA; Ronald, J; Desai, H; Dai, R; Ziegler, C; Nelson, RC; Wildman-Tobriner, B
Published in: Abdom Radiol (NY)
August 2022

PURPOSE: To compare clinical efficacy, subjective radiologist preference, and complication rates for two different core biopsy needles, the Achieve® and Marquee®. METHODS: Retrospective review included consecutive patients who underwent 18 gauge non-targeted core liver biopsy, 30 with Achieve® (Merit Medical) and 30 with Marquee® (BD Bard) Pathologist (blinded to needle type) reviewed specimen total length, maximum width, and portal triad count. Sixteen radiologists subjectively rated (1 to 5(best)) each needle for cocking, firing, recoil, chamber exposure, handling, and overall. A medical records search of all (targeted and non-targeted) core liver biopsies 1/1/17-9/30/2020 compared rates of major (requiring transfusion and/or embolization) and minor (self-limited bleeding) hemorrhagic complications. Comparison between needle types was performed using t-test. RESULTS: For Achieve® and Marquee® needles, the respective mean (SD) for total length(mm) was 29.7(7.0) and 31.9(4.6), p = 0.1; max width(mm) was 0.78(0.1) and 0.85(0.1), p < 0.01; and number of portal triads was 15.3(5.3) and 17.3(5.3), p = 0.2. Radiologists subjectively preferred the Marquee® for several measures including cocking, chamber exposure, and overall (p < 0.02 for each), while the needles were rated similarly for firing, recoil, and handling. Review of 800 cases showed no difference in major (1.0% Achieve®, 1.9% Marquee®, p = 0.5) or minor (1.5% Achieve®, 0.5% Marquee®, p = 0.3) rates of hemorrhagic complications. CONCLUSION: Liver biopsy specimens were significantly wider with Marquee® compared to Achieve®. Radiologists preferred the Marquee® for multiple tactile measures, while the major complication rate was not significantly different. While both needles have a similar side-notch design, the Marquee® needle demonstrates better sample quality and higher user preference, without compromising safety.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Abdom Radiol (NY)

DOI

EISSN

2366-0058

Publication Date

August 2022

Volume

47

Issue

8

Start / End Page

2632 / 2639

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Needles
  • Liver
  • Humans
  • Biopsy, Large-Core Needle
  • Biopsy
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Ho, L. M., Pendse, A. A., Ronald, J., Desai, H., Dai, R., Ziegler, C., … Wildman-Tobriner, B. (2022). Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®. Abdom Radiol (NY), 47(8), 2632–2639. https://doi.org/10.1007/s00261-021-03187-5
Ho, Lisa M., Avani A. Pendse, James Ronald, Hemant Desai, Rui Dai, Cole Ziegler, Rendon C. Nelson, and Benjamin Wildman-Tobriner. “Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®.Abdom Radiol (NY) 47, no. 8 (August 2022): 2632–39. https://doi.org/10.1007/s00261-021-03187-5.
Ho LM, Pendse AA, Ronald J, Desai H, Dai R, Ziegler C, et al. Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®. Abdom Radiol (NY). 2022 Aug;47(8):2632–9.
Ho, Lisa M., et al. “Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®.Abdom Radiol (NY), vol. 47, no. 8, Aug. 2022, pp. 2632–39. Pubmed, doi:10.1007/s00261-021-03187-5.
Ho LM, Pendse AA, Ronald J, Desai H, Dai R, Ziegler C, Nelson RC, Wildman-Tobriner B. Comparison of clinical efficacy, subjective user experience, and safety for two different core biopsy needles, the Achieve® and Marquee®. Abdom Radiol (NY). 2022 Aug;47(8):2632–2639.
Journal cover image

Published In

Abdom Radiol (NY)

DOI

EISSN

2366-0058

Publication Date

August 2022

Volume

47

Issue

8

Start / End Page

2632 / 2639

Location

United States

Related Subject Headings

  • Treatment Outcome
  • Needles
  • Liver
  • Humans
  • Biopsy, Large-Core Needle
  • Biopsy