Skip to main content

Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields.

Publication ,  Journal Article
Jacqmin, DJ; Miller, JR; Barraclough, BA; Labby, ZE
Published in: J Appl Clin Med Phys
August 2022

PURPOSE: The purpose of this work is to evaluate the Standard Imaging Exradin W2 plastic scintillation detector (W2) for use in the types of fields used for stereotactic radiosurgery. METHODS: Prior to testing the W2 in small fields, the W2 was evaluated in standard large field conditions to ensure good detector performance. These tests included energy dependence, short-term repeatability, dose-response linearity, angular dependence, temperature dependence, and dose rate dependence. Next, scan settings and calibration of the W2 were optimized to ensure high quality data acquisition. Profiles of small fields shaped by cones and multi-leaf collimator (MLCs) were measured using the W2 and IBA RAZOR diode in a scanning water tank. Output factors for cones (4-17.5 mm) and MLC fields (1, 2, 3 cm) were acquired with both detectors. Finally, the dose at isocenter for seven radiosurgery plans was measured with the W2 detector. RESULTS: W2 exhibited acceptable warm-up behavior, short-term reproducibility, axial angular dependence, dose-rate linearity, and dose linearity. The detector exhibits a dependence upon energy, polar angle, and temperature. Scanning measurements taken with the W2 and RAZOR were in good agreement, with full-width half-maximum and penumbra widths agreeing to within 0.1 mm. The output factors measured by the W2 and RAZOR exhibited a maximum difference of 1.8%. For the seven point-dose measurements of radiosurgery plans, the W2 agreed well with our treatment planning system with a maximum deviation of 2.2%. The Čerenkov light ratio calibration method did not significantly impact the measurement of relative profiles, output factors, or point dose measurements. CONCLUSION: The W2 demonstrated dosimetric characteristics that are suitable for radiosurgery field measurements. The detector agreed well with the RAZOR diode for output factors and scanned profiles and showed good agreement with the treatment planning system in measurements of clinical treatment plans.

Duke Scholars

Published In

J Appl Clin Med Phys

DOI

EISSN

1526-9914

Publication Date

August 2022

Volume

23

Issue

8

Start / End Page

e13728

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Radiosurgery
  • Radiometry
  • Plastics
  • Nuclear Medicine & Medical Imaging
  • Humans
  • Calibration
  • 5105 Medical and biological physics
  • 3208 Medical physiology
  • 1116 Medical Physiology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Jacqmin, D. J., Miller, J. R., Barraclough, B. A., & Labby, Z. E. (2022). Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields. J Appl Clin Med Phys, 23(8), e13728. https://doi.org/10.1002/acm2.13728
Jacqmin, Dustin J., Jessica R. Miller, Brendan A. Barraclough, and Zacariah E. Labby. “Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields.J Appl Clin Med Phys 23, no. 8 (August 2022): e13728. https://doi.org/10.1002/acm2.13728.
Jacqmin DJ, Miller JR, Barraclough BA, Labby ZE. Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields. J Appl Clin Med Phys. 2022 Aug;23(8):e13728.
Jacqmin, Dustin J., et al. “Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields.J Appl Clin Med Phys, vol. 23, no. 8, Aug. 2022, p. e13728. Pubmed, doi:10.1002/acm2.13728.
Jacqmin DJ, Miller JR, Barraclough BA, Labby ZE. Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields. J Appl Clin Med Phys. 2022 Aug;23(8):e13728.

Published In

J Appl Clin Med Phys

DOI

EISSN

1526-9914

Publication Date

August 2022

Volume

23

Issue

8

Start / End Page

e13728

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Radiosurgery
  • Radiometry
  • Plastics
  • Nuclear Medicine & Medical Imaging
  • Humans
  • Calibration
  • 5105 Medical and biological physics
  • 3208 Medical physiology
  • 1116 Medical Physiology