An unusual case of pulmonary artery catheter malfunction.

Publication ,  Journal Article
Manecke, GR; Brown, JC; Landau, AA; Kapelanski, DP; St Laurent, CM; Auger, WR
Published in: Anesth Analg
August 2002

Sudden loss of thermodilution or temperature-monitoring capabilities of an indwelling pulmonary artery catheter may indicate significant damage to the catheter, possibly leading to electrical hazard and infection risk. Blood appearing at the electrical connection port confirms the diagnosis. The catheter, if easily removed, should be replaced as soon as possible.

Duke Scholars

Published In

Anesth Analg

DOI

ISSN

0003-2999

Publication Date

August 2002

Volume

95

Issue

2

Start / End Page

302 / 304

Location

United States

Related Subject Headings

  • Thermodilution
  • Monitoring, Intraoperative
  • Middle Aged
  • Hypertension, Pulmonary
  • Humans
  • Heart Septal Defects, Atrial
  • Female
  • Equipment Failure
  • Endarterectomy
  • Catheterization, Swan-Ganz
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Manecke, G. R., Brown, J. C., Landau, A. A., Kapelanski, D. P., St Laurent, C. M., & Auger, W. R. (2002). An unusual case of pulmonary artery catheter malfunction. Anesth Analg, 95(2), 302–304. https://doi.org/10.1097/00000539-200208000-00008
Manecke, Gerard R., Jason C. Brown, Allan A. Landau, David P. Kapelanski, Carla M. St Laurent, and William R. Auger. “An unusual case of pulmonary artery catheter malfunction.Anesth Analg 95, no. 2 (August 2002): 302–4. https://doi.org/10.1097/00000539-200208000-00008.
Manecke GR, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002 Aug;95(2):302–4.
Manecke, Gerard R., et al. “An unusual case of pulmonary artery catheter malfunction.Anesth Analg, vol. 95, no. 2, Aug. 2002, pp. 302–04. Pubmed, doi:10.1097/00000539-200208000-00008.
Manecke GR, Brown JC, Landau AA, Kapelanski DP, St Laurent CM, Auger WR. An unusual case of pulmonary artery catheter malfunction. Anesth Analg. 2002 Aug;95(2):302–304.

Published In

Anesth Analg

DOI

ISSN

0003-2999

Publication Date

August 2002

Volume

95

Issue

2

Start / End Page

302 / 304

Location

United States

Related Subject Headings

  • Thermodilution
  • Monitoring, Intraoperative
  • Middle Aged
  • Hypertension, Pulmonary
  • Humans
  • Heart Septal Defects, Atrial
  • Female
  • Equipment Failure
  • Endarterectomy
  • Catheterization, Swan-Ganz