Pneumatic transport exacerbates interference of room air contamination in blood gas samples.

Publication ,  Journal Article
Astles, JR; Lubarsky, D; Loun, B; Sedor, FA; Toffaletti, JG
Published in: Arch Pathol Lab Med
July 1996

OBJECTIVE: To characterize and control the potential interference to P(O2) determinations when blood contaminated with air is sent via a pneumatic tube system (PTS). DESIGN: Both tonometered blood at P(O2)s of 65, 75, 142, and 339 mm Hg and arterial blood gas samples from patients with baseline P(O2)s from 70 to 400 mm Hg were analyzed for P(O2) to determine possible effects of air contamination from PTS transport. SETTING: A large teaching hospital in which a variety of personnel routinely send samples to the laboratory by PTS transport. PATIENTS: Twenty patients under anesthesia for elective surgery and 21 patients in an intensive care unit who had a wide range of P(O2)s. Several additional patients with a preexisting lung pathology likely to cause hypoxemia were selected to provide samples with low P(O2)s. MAIN OUTCOME MEASURES: Measurement of bias in P(O2) between samples sent via PTS and samples walked to the laboratory. RESULTS: Interference from air contamination was worse after PTS transport compared with manual transport of the specimen. Over a wide range, the P(O2) in specimens after PTS transport tended toward 160 mm Hg. Samples from hypoxemic patients were prone to errors in P(O2) that could have resulted in clinical misinterpretation; 5 of 10 samples with a baseline P(O2) less than 85 mm Hg had increases of 10 mm Hg or more when contaminated with air. Cooling samples with high P(O2)s minimized changes to P(O2), probably by increasing the solubility of oxygen. Mechanical buffering by various liners used in the carriers did little to alleviate the interference. Decreasing the speed of pneumatic transport by 50% lessened the effect on P(O2). CONCLUSION: Interference can be minimized by carefully purging samples of all air bubbles using the following protocol: invert syringe to check for air bubbles, then retap and reexpel bubbles if necessary. Personnel that collect and send blood gas samples via PTS should be educated about the problem of interference. Modifications both to pneumatic sample transport systems and to blood gas syringes should be investigated to minimize the effect.

Duke Scholars

Published In

Arch Pathol Lab Med

ISSN

0003-9985

Publication Date

July 1996

Volume

120

Issue

7

Start / End Page

642 / 647

Location

United States

Related Subject Headings

  • Transportation
  • Temperature
  • Reproducibility of Results
  • Pathology
  • Oxygen
  • Oximetry
  • Hypoxia
  • Humans
  • Equipment Contamination
  • Blood Specimen Collection
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Astles, J. R., Lubarsky, D., Loun, B., Sedor, F. A., & Toffaletti, J. G. (1996). Pneumatic transport exacerbates interference of room air contamination in blood gas samples. Arch Pathol Lab Med, 120(7), 642–647.
Astles, J. R., D. Lubarsky, B. Loun, F. A. Sedor, and J. G. Toffaletti. “Pneumatic transport exacerbates interference of room air contamination in blood gas samples.Arch Pathol Lab Med 120, no. 7 (July 1996): 642–47.
Astles JR, Lubarsky D, Loun B, Sedor FA, Toffaletti JG. Pneumatic transport exacerbates interference of room air contamination in blood gas samples. Arch Pathol Lab Med. 1996 Jul;120(7):642–7.
Astles, J. R., et al. “Pneumatic transport exacerbates interference of room air contamination in blood gas samples.Arch Pathol Lab Med, vol. 120, no. 7, July 1996, pp. 642–47.
Astles JR, Lubarsky D, Loun B, Sedor FA, Toffaletti JG. Pneumatic transport exacerbates interference of room air contamination in blood gas samples. Arch Pathol Lab Med. 1996 Jul;120(7):642–647.

Published In

Arch Pathol Lab Med

ISSN

0003-9985

Publication Date

July 1996

Volume

120

Issue

7

Start / End Page

642 / 647

Location

United States

Related Subject Headings

  • Transportation
  • Temperature
  • Reproducibility of Results
  • Pathology
  • Oxygen
  • Oximetry
  • Hypoxia
  • Humans
  • Equipment Contamination
  • Blood Specimen Collection