Effects of characteristic x-rays on assay of I-123 by dose calibrator.

Publication ,  Journal Article
Harris, CC; Jaszczak, RJ; Greer, KL; Briner, WH; Coleman, RE
Published in: J Nucl Med
December 1984

Large differences in dose-calibrator readings are obtained if "high-purity" I-123 is assayed in different containers. Large correction factors are necessary for assaying another isotope of iodine, I-125, in a dose calibrator, because of absorption of the low-energy (28.4-keV, weighted mean) emissions. We found that up to 70% of the dose-calibrator response to I-123 can be due to characteristic x-rays with energies exactly the same as those emitted by I-125, and that dose-calibrator response to I-123 is also strongly affected by the absorption properties of the vial. An appropriate method to define I-123 activity uses a gamma camera with a medium-energy collimator to establish correction factors for dose-calibrator assay of I-123 in different containers. Correction factors for a plastic syringe and a thick-wall glass vial, were determined using this method. Measurement of I-123 activity in a copper absorber will eliminate the response to x-rays, and the gamma camera is useful in establishing the necessary correction factors.

Duke Scholars

Published In

J Nucl Med

ISSN

0161-5505

Publication Date

December 1984

Volume

25

Issue

12

Start / End Page

1367 / 1370

Location

United States

Related Subject Headings

  • X-Rays
  • Radioactivity
  • Nuclear Medicine & Medical Imaging
  • Iodine Radioisotopes
  • Calibration
  • 3202 Clinical sciences
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Harris, C. C., Jaszczak, R. J., Greer, K. L., Briner, W. H., & Coleman, R. E. (1984). Effects of characteristic x-rays on assay of I-123 by dose calibrator. J Nucl Med, 25(12), 1367–1370.
Harris, C. C., R. J. Jaszczak, K. L. Greer, W. H. Briner, and R. E. Coleman. “Effects of characteristic x-rays on assay of I-123 by dose calibrator.J Nucl Med 25, no. 12 (December 1984): 1367–70.
Harris CC, Jaszczak RJ, Greer KL, Briner WH, Coleman RE. Effects of characteristic x-rays on assay of I-123 by dose calibrator. J Nucl Med. 1984 Dec;25(12):1367–70.
Harris, C. C., et al. “Effects of characteristic x-rays on assay of I-123 by dose calibrator.J Nucl Med, vol. 25, no. 12, Dec. 1984, pp. 1367–70.
Harris CC, Jaszczak RJ, Greer KL, Briner WH, Coleman RE. Effects of characteristic x-rays on assay of I-123 by dose calibrator. J Nucl Med. 1984 Dec;25(12):1367–1370.

Published In

J Nucl Med

ISSN

0161-5505

Publication Date

December 1984

Volume

25

Issue

12

Start / End Page

1367 / 1370

Location

United States

Related Subject Headings

  • X-Rays
  • Radioactivity
  • Nuclear Medicine & Medical Imaging
  • Iodine Radioisotopes
  • Calibration
  • 3202 Clinical sciences
  • 1103 Clinical Sciences