An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry.

Publication ,  Journal Article
Oldham, M; Baustert, I; Lord, C; Smith, TA; McJury, M; Warrington, AP; Leach, MO; Webb, S
Published in: Phys Med Biol
May 1998

BANG-gel dosimetry offers the potential for measuring the dose delivered by a radiotherapy treatment technique, in three dimensions, with high spatial resolution and good accuracy. The ability to measure comprehensively a 3D dose distribution is a major advantage of the gel dosimeter over conventional planar and point-based dosimeter devices, particularly when applied to the verification of complex dose distributions characteristic of intensity-modulated radiotherapy (IMRT). In this paper an in-house manufactured BANG-gel dosimeter was applied to study the dose distributions of two irradiation experiments for which the distributions were known: (i) a dosimetrically simple parallel-opposed irradiation, and (ii) a more complex nine-field 'static tomotherapy' intensity-modulated irradiation delivered with the Nomos MIMiC. The uniform distribution in (i) allowed a study of the magnetic resonance (MR) imaging parameters to achieve an optimal trade-off between noise and image resolution (optimum image resolution for the Siemens 1.5T Vision system was determined to be approximately 0.8 mm2 with a slice thickness of 2 mm). The spatial uniformity of gel sensitivity to radiation was found to depend strongly on the presence of oxygen, which must be eliminated for the gel dosimeter to be of use. The gel dosimeter was found to agree well with predicted dose distributions and accurately measured the steep penumbral fall-off of dose, even after many days, proving its potential for the verification of IMRT distributions. In the nine-field IMRT delivery (ii) the predicted dose was computed by both an in-house 'component-delivery' dose algorithm and the Peacock planning-system dose algorithm. Good agreement was found between the two algorithms despite the latter's omission of the change in penumbral characteristics with aperture-size during delivery, lack of inhomogeneity correction and approximate modelling of leaf leakage. These effects were found to be small for the problem studied. The predicted distribution agreed well with the gel-measured distribution at medium and high doses (50-90% isodose lines) although differences of up to 10% were observed at lower doses (30% isodose line). The gel dosimeter was found to have the potential to verify IMRT distributions but required considerable care to achieve accurate results. Attention was required to achieve uniformity of gel sensitivity (to prevent oxygen contamination), and in the calibration process.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

May 1998

Volume

43

Issue

5

Start / End Page

1113 / 1132

Location

England

Related Subject Headings

  • Water
  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiographic Image Interpretation, Computer-Assisted
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Humans
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Oldham, M., Baustert, I., Lord, C., Smith, T. A., McJury, M., Warrington, A. P., … Webb, S. (1998). An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry. Phys Med Biol, 43(5), 1113–1132. https://doi.org/10.1088/0031-9155/43/5/005
Oldham, M., I. Baustert, C. Lord, T. A. Smith, M. McJury, A. P. Warrington, M. O. Leach, and S. Webb. “An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry.Phys Med Biol 43, no. 5 (May 1998): 1113–32. https://doi.org/10.1088/0031-9155/43/5/005.
Oldham M, Baustert I, Lord C, Smith TA, McJury M, Warrington AP, et al. An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry. Phys Med Biol. 1998 May;43(5):1113–32.
Oldham, M., et al. “An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry.Phys Med Biol, vol. 43, no. 5, May 1998, pp. 1113–32. Pubmed, doi:10.1088/0031-9155/43/5/005.
Oldham M, Baustert I, Lord C, Smith TA, McJury M, Warrington AP, Leach MO, Webb S. An investigation into the dosimetry of a nine-field tomotherapy irradiation using BANG-gel dosimetry. Phys Med Biol. 1998 May;43(5):1113–1132.
Journal cover image

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

May 1998

Volume

43

Issue

5

Start / End Page

1113 / 1132

Location

England

Related Subject Headings

  • Water
  • Tomography, X-Ray Computed
  • Sensitivity and Specificity
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiographic Image Interpretation, Computer-Assisted
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • Humans