Skip to main content

Dialysate and blood flow dependence of diffusive solute clearance during CVVHD.

Publication ,  Journal Article
Relton, S; Greenberg, A; Palevsky, PM
Published in: ASAIO J
1992

To define the flow dependence of diffusive solute clearance (Kd) during continuous venovenous hemodiafiltration (CVVHD), urea, creatinine, and phosphate clearance were measured during CVVHD, while ultrafiltration was minimized. Studies were performed using AN69, polyacrylonitrile (PAN), and polyamide (PA) hollow fiber hemofilters. To assess the dialysate flow dependence of Kd, clearances were measured at constant blood flow (QB), while dialysate inflow (QDi) was increased incrementally from 8.3 to 33.3 ml/min. Kd increased linearly for all three solutes using all membranes. To assess its blood flow dependence, Kd was measured at blood flows of 50, 75, 100, and 150 ml/min, while QDi was held constant at 33.3 ml/min. The PA membrane showed no flow dependence; for the other membranes, Kd was flow independent for QB > 100 ml/min. At all values of QB and QDi, clearances for the AN69 and PAN membranes were two to three times that of the PA membrane (p < 0.01). Lactate uptake increased with increasing QDi for all three membranes, but did not result in blood lactate accumulation. It is concluded that: 1) small molecule clearance is dialysate flow limited during CVVHD; 2) increasing QB beyond 100 ml/min is unnecessary for maintaining solute clearance; and 3) the PA membrane is unsuitable for CVVHD because of its low diffusive solute clearance.

Duke Scholars

Published In

ASAIO J

DOI

ISSN

1058-2916

Publication Date

1992

Volume

38

Issue

3

Start / End Page

M691 / M696

Location

United States

Related Subject Headings

  • Urea
  • Phosphates
  • Lactic Acid
  • Lactates
  • Kinetics
  • Kidney Failure, Chronic
  • Humans
  • Hemofiltration
  • Dialysis Solutions
  • Creatinine
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Relton, S., Greenberg, A., & Palevsky, P. M. (1992). Dialysate and blood flow dependence of diffusive solute clearance during CVVHD. ASAIO J, 38(3), M691–M696. https://doi.org/10.1097/00002480-199207000-00127
Relton, S., A. Greenberg, and P. M. Palevsky. “Dialysate and blood flow dependence of diffusive solute clearance during CVVHD.ASAIO J 38, no. 3 (1992): M691–96. https://doi.org/10.1097/00002480-199207000-00127.
Relton S, Greenberg A, Palevsky PM. Dialysate and blood flow dependence of diffusive solute clearance during CVVHD. ASAIO J. 1992;38(3):M691–6.
Relton, S., et al. “Dialysate and blood flow dependence of diffusive solute clearance during CVVHD.ASAIO J, vol. 38, no. 3, 1992, pp. M691–96. Pubmed, doi:10.1097/00002480-199207000-00127.
Relton S, Greenberg A, Palevsky PM. Dialysate and blood flow dependence of diffusive solute clearance during CVVHD. ASAIO J. 1992;38(3):M691–M696.

Published In

ASAIO J

DOI

ISSN

1058-2916

Publication Date

1992

Volume

38

Issue

3

Start / End Page

M691 / M696

Location

United States

Related Subject Headings

  • Urea
  • Phosphates
  • Lactic Acid
  • Lactates
  • Kinetics
  • Kidney Failure, Chronic
  • Humans
  • Hemofiltration
  • Dialysis Solutions
  • Creatinine