Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette.

Publication ,  Journal Article
Floyd, CE; Chotas, HG; Ravin, CE
Published in: Invest Radiol
September 1994

RATIONALE AND OBJECTIVES: The scatter-reduction properties of a new infinity-focused grid incorporated into a radiographic imaging cassette were analyzed. METHODS: A polystyrene chest phantom was imaged using the cassette and bedside radiographic procedure. Scatter fractions were measured using a beam-stop technique at several anatomically equivalent locations. The performance of this cassette also was evaluated as a function of the orientation angle. RESULTS: The gridded cassette provided a decrease in scatter fraction from 61% to 49% in the lung and from 87% to 76% in the mediastinum. To obtain equivalent film density when the grid was used, the exposure was increased by a factor of 3.0. While there was a decrease in scatter clean-up as the film cassette was tilted from the perpendicular, the grid performed well out to an angle of 10 degrees. CONCLUSION: The gridded cassette for portable radiography provides scatter reduction with little sensitivity to alignment. The availability of this device could improve scatter rejection, and therefore, contrast in portable radiographic imaging.

Duke Scholars

Published In

Invest Radiol

DOI

ISSN

0020-9996

Publication Date

September 1994

Volume

29

Issue

9

Start / End Page

852 / 855

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Technology, Radiologic
  • Scattering, Radiation
  • Radiography, Thoracic
  • Radiography
  • Nuclear Medicine & Medical Imaging
  • Humans
  • 3202 Clinical sciences
  • 1103 Clinical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Floyd, C. E., Chotas, H. G., & Ravin, C. E. (1994). Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette. Invest Radiol, 29(9), 852–855. https://doi.org/10.1097/00004424-199409000-00010
Floyd, C. E., H. G. Chotas, and C. E. Ravin. “Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette.Invest Radiol 29, no. 9 (September 1994): 852–55. https://doi.org/10.1097/00004424-199409000-00010.
Floyd CE, Chotas HG, Ravin CE. Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette. Invest Radiol. 1994 Sep;29(9):852–5.
Floyd, C. E., et al. “Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette.Invest Radiol, vol. 29, no. 9, Sept. 1994, pp. 852–55. Pubmed, doi:10.1097/00004424-199409000-00010.
Floyd CE, Chotas HG, Ravin CE. Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette. Invest Radiol. 1994 Sep;29(9):852–855.

Published In

Invest Radiol

DOI

ISSN

0020-9996

Publication Date

September 1994

Volume

29

Issue

9

Start / End Page

852 / 855

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Technology, Radiologic
  • Scattering, Radiation
  • Radiography, Thoracic
  • Radiography
  • Nuclear Medicine & Medical Imaging
  • Humans
  • 3202 Clinical sciences
  • 1103 Clinical Sciences