Hydrologic research is a very demanding application of fiber-optic distributed temperature sensing (DTS) in terms of precision, accuracy and calibration. The physics behind the most frequently used DTS instruments are considered as they apply to four calibration methods for single-ended DTS installations. The new methods presented are more accurate than the instrument-calibrated data, achieving accuracies on the order of tenths of a degree root mean square error (RMSE) and mean bias. Effects of localized non-uniformities that violate the assumptions of single-ended calibration data are explored and quantified. Experimental design considerations such as selection of integration times or selection of the length of the reference sections are discussed, and the impacts of these considerations on calibrated temperatures are explored in two case studies.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Sensors (Basel, Switzerland)

DOI

EISSN

1424-8220

ISSN

1424-8220

Publication Date

January 2011

Volume

11

Issue

11

Start / End Page

10859 / 10879

Related Subject Headings

  • Trees
  • Thermometers
  • Temperature
  • Spectrum Analysis, Raman
  • Remote Sensing Technology
  • Ponds
  • Fiber Optic Technology
  • Ecosystem
  • Calibration
  • Analytical Chemistry
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Hausner, M. B., Suárez, F., Glander, K. E., van de Giesen, N., Selker, J. S., & Tyler, S. W. (2011). Calibrating single-ended fiber-optic Raman spectra distributed temperature sensing data. Sensors (Basel, Switzerland), 11(11), 10859–10879. https://doi.org/10.3390/s111110859
Hausner, Mark B., Francisco Suárez, Kenneth E. Glander, Nick van de Giesen, John S. Selker, and Scott W. Tyler. “Calibrating single-ended fiber-optic Raman spectra distributed temperature sensing data.Sensors (Basel, Switzerland) 11, no. 11 (January 2011): 10859–79. https://doi.org/10.3390/s111110859.
Hausner MB, Suárez F, Glander KE, van de Giesen N, Selker JS, Tyler SW. Calibrating single-ended fiber-optic Raman spectra distributed temperature sensing data. Sensors (Basel, Switzerland). 2011 Jan;11(11):10859–79.
Hausner, Mark B., et al. “Calibrating single-ended fiber-optic Raman spectra distributed temperature sensing data.Sensors (Basel, Switzerland), vol. 11, no. 11, Jan. 2011, pp. 10859–79. Epmc, doi:10.3390/s111110859.
Hausner MB, Suárez F, Glander KE, van de Giesen N, Selker JS, Tyler SW. Calibrating single-ended fiber-optic Raman spectra distributed temperature sensing data. Sensors (Basel, Switzerland). 2011 Jan;11(11):10859–10879.

Published In

Sensors (Basel, Switzerland)

DOI

EISSN

1424-8220

ISSN

1424-8220

Publication Date

January 2011

Volume

11

Issue

11

Start / End Page

10859 / 10879

Related Subject Headings

  • Trees
  • Thermometers
  • Temperature
  • Spectrum Analysis, Raman
  • Remote Sensing Technology
  • Ponds
  • Fiber Optic Technology
  • Ecosystem
  • Calibration
  • Analytical Chemistry