Skip to main content

Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol.

Publication ,  Journal Article
Bashir, MR; Dale, BM; Gupta, RT; Horvath, JJ; Boll, DT; Merkle, EM
Published in: Invest Radiol
September 2012

PURPOSE: The aim of this study was to examine the effect of minimizing prescan adjustments on table time and image quality in magnetic resonance imaging (MRI) of the liver. MATERIALS AND METHODS: This prospective Health Insurance Portability and Accountability Act-compliant study was approved by the institutional review board, and written informed consent was obtained. Ten volunteers were imaged twice using a noncontrast liver MRI protocol consisting of a total of 10 pulse sequences, once with a standard protocol and once with a fixed table position/minimized prescan adjustment protocol (in random order). Total examination time was evaluated according to a Lean Six Sigma framework. Quantitative sequences, including diffusion-weighted imaging with apparent diffusion coefficient, multi-echo Dixon fat percentage, and the transverse relaxation time, were evaluated and compared between the two protocols. Two experienced readers, blinded to the protocol used, compared image quality between the two protocols. RESULTS: The average number of prescan adjustment steps per examination was reduced from 58.0 to 22.1 using the minimal shimming protocol compared with the normal shimming protocol (P < 0.001). Mean business value added time (scan preparatory time) was reduced by 58% (3 minutes 3 seconds vs 7 minutes 13 seconds), whereas mean total examination time was 20% lower (18 minutes 13 seconds vs 22 minutes 48 seconds, P < 0.001). Quantitative measures obtained using the two protocols were equivalent, and neither reader detected a significant difference in subjective image quality. CONCLUSION: Fixing table position minimizes prescan adjustments and reduces total table time in liver MRI without adversely affecting image quality.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Invest Radiol

DOI

EISSN

1536-0210

Publication Date

September 2012

Volume

47

Issue

9

Start / End Page

524 / 529

Location

United States

Related Subject Headings

  • Prospective Studies
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Magnetic Resonance Imaging
  • Liver
  • Image Enhancement
  • Humans
  • Female
  • Clinical Protocols
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Bashir, M. R., Dale, B. M., Gupta, R. T., Horvath, J. J., Boll, D. T., & Merkle, E. M. (2012). Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol. Invest Radiol, 47(9), 524–529. https://doi.org/10.1097/RLI.0b013e31825a8e5b
Bashir, Mustafa R., Brian M. Dale, Rajan T. Gupta, Jeffrey J. Horvath, Daniel T. Boll, and Elmar M. Merkle. “Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol.Invest Radiol 47, no. 9 (September 2012): 524–29. https://doi.org/10.1097/RLI.0b013e31825a8e5b.
Bashir MR, Dale BM, Gupta RT, Horvath JJ, Boll DT, Merkle EM. Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol. Invest Radiol. 2012 Sep;47(9):524–9.
Bashir, Mustafa R., et al. “Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol.Invest Radiol, vol. 47, no. 9, Sept. 2012, pp. 524–29. Pubmed, doi:10.1097/RLI.0b013e31825a8e5b.
Bashir MR, Dale BM, Gupta RT, Horvath JJ, Boll DT, Merkle EM. Gradient shimming during magnetic resonance imaging of the liver: comparison of a standard protocol versus a novel reduced protocol. Invest Radiol. 2012 Sep;47(9):524–529.

Published In

Invest Radiol

DOI

EISSN

1536-0210

Publication Date

September 2012

Volume

47

Issue

9

Start / End Page

524 / 529

Location

United States

Related Subject Headings

  • Prospective Studies
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Magnetic Resonance Imaging
  • Liver
  • Image Enhancement
  • Humans
  • Female
  • Clinical Protocols