Late safety, efficacy, and cost-effectiveness of a zotarolimus-eluting stent compared with a paclitaxel-eluting stent in patients with de novo coronary lesions: 2-year follow-up from the ENDEAVOR IV trial (Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions).

Publication ,  Journal Article
Leon, MB; Kandzari, DE; Eisenstein, EL; Anstrom, KJ; Mauri, L; Cutlip, DE; Nikolsky, E; O'Shaughnessy, C; Overlie, PA; Kirtane, AJ; Solomon, SL ...
Published in: JACC Cardiovasc Interv
December 2009

OBJECTIVES: The aim of this study was to assess, after 2 years of follow-up, the safety, efficacy, and cost-effectiveness of a zotarolimus-eluting stent (ZES) compared with a paclitaxel-eluting stent (PES) in patients with native coronary lesions. BACKGROUND: Early drug-eluting stents were associated with a small but significant incidence of very late stent thrombosis (VLST), occurring >1 year after the index procedure. The ZES has shown encouraging results in clinical trials. METHODS: The ENDEAVOR IV trial (Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions), a randomized (1:1), single-blind, controlled trial (n = 1,548) compared ZES versus PES in patients with single de novo coronary lesions. Two-year follow-up was obtained in 96.0% of ZES and 95.4% of PES patients. The primary end point was target vessel failure (TVF), and safety end points included Academic Research Consortium-defined stent thrombosis. Economic end points analyzed included quality-adjusted survival, medical costs, and relative cost-effectiveness of ZES and PES. RESULTS: The TVF at 2 years was similar in ZES and PES patients (11.1% vs. 13.1%, p = 0.232). There were fewer myocardial infarctions (MIs) in ZES patients (p = 0.022), due to fewer periprocedural non-Q-wave MIs and fewer late MIs between 1 and 2 years. Late MIs were associated with increased VLST (PES: 6 vs. ZES: 1; p = 0.069). Target lesion revascularization was similar comparing ZES with PES (5.9% vs. 4.6%; p = 0.295), especially in patients without planned angiographic follow-up (5.2% vs. 4.9%; p = 0.896). The cost-effectiveness of ZES and PES was similar. CONCLUSIONS: After 2 years of follow-up, ZES demonstrated efficacy and cost-effectiveness comparable to PES, with fewer MIs and a trend toward less VLST. (The ENDEAVOR IV Clinical Trial: A Trial of a Coronary Stent System in Coronary Artery Lesions; NCT00217269).

Duke Scholars

Published In

JACC Cardiovasc Interv

DOI

EISSN

1876-7605

Publication Date

December 2009

Volume

2

Issue

12

Start / End Page

1208 / 1218

Location

United States

Related Subject Headings

  • United States
  • Treatment Outcome
  • Time Factors
  • Sirolimus
  • Single-Blind Method
  • Risk Assessment
  • Quality-Adjusted Life Years
  • Quality of Life
  • Prosthesis Design
  • Prospective Studies
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Journal cover image

Published In

JACC Cardiovasc Interv

DOI

EISSN

1876-7605

Publication Date

December 2009

Volume

2

Issue

12

Start / End Page

1208 / 1218

Location

United States

Related Subject Headings

  • United States
  • Treatment Outcome
  • Time Factors
  • Sirolimus
  • Single-Blind Method
  • Risk Assessment
  • Quality-Adjusted Life Years
  • Quality of Life
  • Prosthesis Design
  • Prospective Studies