Skip to main content

Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram

Publication ,  Journal Article
Nolle, FM; Bowser, RW; Badura, FK; Catlett, JM; Guadapati, RR; Hee, TT; Mooss, AN; Sketch, MH
Published in: Computers in Cardiology
September 1, 1988

The authors evaluate a frequency-domain algorithm for use by the CREI-GARD computerized arrhythmia system (F. M. Nolle et al., 1986) to detect ventricular fibrillation (VF) in the surface ECG. This was accomplished using (1) The American Heart Association (AHA) database of 80 ECG tape recordings (channel 1), including 10 recordings with 11 documented VF episodes, and (2) The CREI-GARD database of 35 single-channel recordings of 35 documented VF episodes. The recordings were analyzed using sequential 4.096-s segments. All 46 episodes of VF were correctly identified. The average time between the onset of VF and the first detection was 3 s (range 0 to 9 s). VF episodes occurring in segments in association with ventricular tachycardia (VT) were all correctly identified. There are 19 false VF detections during other episodes of VT out of a total of 684 VT episodes (2.8%). There were 5 false VF detections during other rhythms. None occurred during artifact. The algorithm is described in detail.

Duke Scholars

Published In

Computers in Cardiology

ISSN

0276-6574

Publication Date

September 1, 1988

Start / End Page

337 / 340

Related Subject Headings

  • Cardiovascular System & Hematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nolle, F. M., Bowser, R. W., Badura, F. K., Catlett, J. M., Guadapati, R. R., Hee, T. T., … Sketch, M. H. (1988). Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram. Computers in Cardiology, 337–340.
Nolle, F. M., R. W. Bowser, F. K. Badura, J. M. Catlett, R. R. Guadapati, T. T. Hee, A. N. Mooss, and M. H. Sketch. “Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram.” Computers in Cardiology, September 1, 1988, 337–40.
Nolle FM, Bowser RW, Badura FK, Catlett JM, Guadapati RR, Hee TT, et al. Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram. Computers in Cardiology. 1988 Sep 1;337–40.
Nolle, F. M., et al. “Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram.” Computers in Cardiology, Sept. 1988, pp. 337–40.
Nolle FM, Bowser RW, Badura FK, Catlett JM, Guadapati RR, Hee TT, Mooss AN, Sketch MH. Evaluation of a frequency-domain algorithm to detect ventricular fibrillation in the surface elctrocardiogram. Computers in Cardiology. 1988 Sep 1;337–340.

Published In

Computers in Cardiology

ISSN

0276-6574

Publication Date

September 1, 1988

Start / End Page

337 / 340

Related Subject Headings

  • Cardiovascular System & Hematology