Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.

Publication ,  Journal Article
McMahon, R; Zhuang, T; Steffey, BA; Song, H; Craciunescu, OI
Published in: J Appl Clin Med Phys
November 15, 2011

Studies have shown that source dwells within Varian's HDR CT/MR compatible ring applicators can deviate from intended positions by several millimeters. Quantifying this offset is an important part of commissioning. The aims of this study were to: 1) determine the reproducibility of the offset, 2) study the interobserver variation in the offset's measurement, and 3) quantify the dosimetric impact of the offset. Offsets were measured for four ring applicators: two 30°, one 45°, and one 60°. Dwell positions were measured five times for each ring to determine the reproducibility of source positioning. Experiments were done to compare two separate source wires, as well as different time points within a single source wire's lifecycle. Data were analyzed by three independent observers. To quantify the dosimetric impact of the offset, a treatment plan was generated using BrachyVision. The dose to point A, and the D(2cc) metric for rectum and bladder were calculated with and without the offset. For the 45° and 60° rings, measured offsets were 3.0 mm and 3.6 mm, respectively. The 30° ring showed substantial variation in distal dwell positions (maximum difference between the five experiments of 2.9 mm). Subsequent testing of a replacement ring showed an offset of 2.4 mm that was more reproducible. Offsets varied less than 1 mm between different source wires, and changed less than 1 mm over the course of a source wire's lifecycle. When comparing observers, the average range in a measurement of a dwell position was 0. 5 mm (σ = 0.2 mm, max 1.3 mm). The offset resulted in dose variations to point A, bladder, and rectum of less than 1%, 2%, and 5%, respectively. Results indicate that Varian rings can show systematic and random offsets of more than 3 mm. Some can be considered defective and should be replaced. Each applicator should be individually commissioned and reproducibility should be confirmed with multiple tests.

Duke Scholars

Published In

J Appl Clin Med Phys

DOI

EISSN

1526-9914

Publication Date

November 15, 2011

Volume

12

Issue

4

Start / End Page

3447

Location

United States

Related Subject Headings

  • Urinary Bladder
  • Tomography, X-Ray Computed
  • Reproducibility of Results
  • Rectum
  • Radiotherapy Planning, Computer-Assisted
  • Observer Variation
  • Nuclear Medicine & Medical Imaging
  • Humans
  • 5105 Medical and biological physics
  • 3208 Medical physiology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
McMahon, R., Zhuang, T., Steffey, B. A., Song, H., & Craciunescu, O. I. (2011). Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. J Appl Clin Med Phys, 12(4), 3447. https://doi.org/10.1120/jacmp.v12i4.3447
McMahon, Ryan, Tingliang Zhuang, Beverly A. Steffey, Haijun Song, and Oana I. Craciunescu. “Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.J Appl Clin Med Phys 12, no. 4 (November 15, 2011): 3447. https://doi.org/10.1120/jacmp.v12i4.3447.
McMahon R, Zhuang T, Steffey BA, Song H, Craciunescu OI. Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. J Appl Clin Med Phys. 2011 Nov 15;12(4):3447.
McMahon, Ryan, et al. “Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.J Appl Clin Med Phys, vol. 12, no. 4, Nov. 2011, p. 3447. Pubmed, doi:10.1120/jacmp.v12i4.3447.
McMahon R, Zhuang T, Steffey BA, Song H, Craciunescu OI. Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets. J Appl Clin Med Phys. 2011 Nov 15;12(4):3447.

Published In

J Appl Clin Med Phys

DOI

EISSN

1526-9914

Publication Date

November 15, 2011

Volume

12

Issue

4

Start / End Page

3447

Location

United States

Related Subject Headings

  • Urinary Bladder
  • Tomography, X-Ray Computed
  • Reproducibility of Results
  • Rectum
  • Radiotherapy Planning, Computer-Assisted
  • Observer Variation
  • Nuclear Medicine & Medical Imaging
  • Humans
  • 5105 Medical and biological physics
  • 3208 Medical physiology