Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators

Publication ,  Journal Article
Peterchev, AV
Published in: 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011
July 20, 2011

We compare half-bridge and full-bridge circuit topologies for controllable pulse parameter transcranial magnetic stimulation (cTMS) devices. The full bridge can generate two more distinct coil voltage levels for enhanced pulse shaping, does not need an active snubber circuit, and requires lower semiconductor switch voltage ratings, at the cost of necessitating four semiconductor switches and associated drivers compared to two in the half bridge. In cTMS devices it may be beneficial to use coils with higher number of turns and, hence, higher inductance than that optimal in conventional TMS, since this reduces the required capacitance, stored energy, coil current, and, potentially, conduction losses and heating. However, increasing the number of coil turns raises the required capacitor voltage which presents practical limits on the number of turns, especially for very brief pulses. © 2011 IEEE.

Duke Scholars

Published In

2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011

DOI

Publication Date

July 20, 2011

Start / End Page

646 / 649
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Peterchev, A. V. (2011). Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators. 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011, 646–649. https://doi.org/10.1109/NER.2011.5910631
Peterchev, A. V. “Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators.” 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011, July 20, 2011, 646–49. https://doi.org/10.1109/NER.2011.5910631.
Peterchev AV. Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators. 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011. 2011 Jul 20;646–9.
Peterchev, A. V. “Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators.” 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011, July 2011, pp. 646–49. Scopus, doi:10.1109/NER.2011.5910631.
Peterchev AV. Circuit topology comparison and design analysis for controllable pulse parameter transcranial magnetic stimulators. 2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011. 2011 Jul 20;646–649.

Published In

2011 5th International IEEE/EMBS Conference on Neural Engineering, NER 2011

DOI

Publication Date

July 20, 2011

Start / End Page

646 / 649