On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter.

Publication ,  Journal Article
Juang, T; Das, S; Adamovics, J; Benning, R; Oldham, M
Published in: Int J Radiat Oncol Biol Phys
October 1, 2013

PURPOSE: To introduce and evaluate a novel deformable 3-dimensional (3D) dosimetry system (Presage-Def/Optical-CT) and its application toward investigating the accuracy of dose deformation in a commercial deformable image registration (DIR) package. METHODS AND MATERIALS: Presage-Def is a new dosimetry material consisting of an elastic polyurethane matrix doped with radiochromic leuco dye. Radiologic and mechanical properties were characterized using standard techniques. Dose-tracking feasibility was evaluated by comparing dose distributions between dosimeters irradiated with and without 27% lateral compression. A checkerboard plan of 5-mm square fields enabled precise measurement of true deformation using 3D dosimetry. Predicted deformation was determined from a commercial DIR algorithm. RESULTS: Presage-Def exhibited a linear dose response with sensitivity of 0.0032 ΔOD/(Gy∙cm). Mass density is 1.02 g/cm(3), and effective atomic number is within 1.5% of water over a broad (0.03-10 MeV) energy range, indicating good water-equivalence. Elastic characteristics were close to that of liver tissue, with Young's modulus of 13.5-887 kPa over a stress range of 0.233-303 kPa, and Poisson's ratio of 0.475 (SE, 0.036). The Presage-Def/Optical-CT system successfully imaged the nondeformed and deformed dose distributions, with isotropic resolution of 1 mm. Comparison with the predicted deformed 3D dose distribution identified inaccuracies in the commercial DIR algorithm. Although external contours were accurately deformed (submillimeter accuracy), volumetric dose deformation was poor. Checkerboard field positioning and dimension errors of up to 9 and 14 mm, respectively, were identified, and the 3D DIR-deformed dose γ passing rate was only γ(3%/3 mm) = 60.0%. CONCLUSIONS: The Presage-Def/Optical-CT system shows strong potential for comprehensive investigation of DIR algorithm accuracy. Substantial errors in a commercial DIR were found in the conditions evaluated. This work highlights the critical importance of careful validation of DIR algorithms before clinical implementation.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Int J Radiat Oncol Biol Phys

DOI

EISSN

1879-355X

Publication Date

October 1, 2013

Volume

87

Issue

2

Start / End Page

414 / 421

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Oncology & Carcinogenesis
  • Liver
  • Imaging, Three-Dimensional
  • Elasticity Imaging Techniques
  • Dose-Response Relationship, Radiation
  • Dose Fractionation, Radiation
  • Algorithms
  • 5105 Medical and biological physics
  • 3407 Theoretical and computational chemistry
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Juang, T., Das, S., Adamovics, J., Benning, R., & Oldham, M. (2013). On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter. Int J Radiat Oncol Biol Phys, 87(2), 414–421. https://doi.org/10.1016/j.ijrobp.2013.05.045
Juang, Titania, Shiva Das, John Adamovics, Ron Benning, and Mark Oldham. “On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter.Int J Radiat Oncol Biol Phys 87, no. 2 (October 1, 2013): 414–21. https://doi.org/10.1016/j.ijrobp.2013.05.045.
Juang T, Das S, Adamovics J, Benning R, Oldham M. On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter. Int J Radiat Oncol Biol Phys. 2013 Oct 1;87(2):414–21.
Juang, Titania, et al. “On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter.Int J Radiat Oncol Biol Phys, vol. 87, no. 2, Oct. 2013, pp. 414–21. Pubmed, doi:10.1016/j.ijrobp.2013.05.045.
Juang T, Das S, Adamovics J, Benning R, Oldham M. On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter. Int J Radiat Oncol Biol Phys. 2013 Oct 1;87(2):414–421.
Journal cover image

Published In

Int J Radiat Oncol Biol Phys

DOI

EISSN

1879-355X

Publication Date

October 1, 2013

Volume

87

Issue

2

Start / End Page

414 / 421

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Oncology & Carcinogenesis
  • Liver
  • Imaging, Three-Dimensional
  • Elasticity Imaging Techniques
  • Dose-Response Relationship, Radiation
  • Dose Fractionation, Radiation
  • Algorithms
  • 5105 Medical and biological physics
  • 3407 Theoretical and computational chemistry