Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy.

Publication ,  Journal Article
Fliotsos, MJ; Deljookorani, S; Dzhaber, D; Chandan, S; Ighani, M; Eghrari, AO
Published in: Cornea
July 2020

We sought to demonstrate the feasibility of a lower-cost, portable method for qualitative and quantitative analysis of the corneal endothelium using a smartphone and slit-lamp biomicroscope.In this study, at a single academic center, we recruited healthy participants to undergo imaging of the corneal endothelium using both a smartphone-based method and a specular microscope. Participants first had their eyes imaged with a CellChek NSP-9900 Specular Microscope (Konan Medical, Inc, Irvine, CA). For image capture on the smartphone, a beam of light approximately 0.2 mm in diameter was directed to the center of the cornea with a slit-lamp biomicroscope to achieve specular reflection. With 40× zoom on the slit-lamp and 4K video mode set on an iPhone 7 Plus held to an ocular, the corneal endothelium was recorded until the hexagonal pattern of cells was identified and the sharpest frame from the video was selected.The videos were analyzed from 14 sets of eyes (average length 2 minutes 40 seconds). The average intraclass correlation coefficient was 0.67 (95% confidence interval, 0.43-0.82). The mean difference between smartphone endothelial cell count and specular endothelial cell count was -209 cells/mm (SD = 483 cells/mm), which did not achieve significance (P = 0.14). A Bland-Altman analysis with simple linear regression showed no proportional bias when comparing the 2 modalities (coefficient = -0.20; t-value = -0.42; P = 0.68).Smartphone specular microscopy is capable of qualitative and quantitative analysis of the corneal endothelium. Further refinement to standardize the light source and automate analysis will increase feasibility.

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Cornea

DOI

EISSN

1536-4798

ISSN

0277-3740

Publication Date

July 2020

Volume

39

Issue

7

Start / End Page

924 / 929

Related Subject Headings

  • Young Adult
  • Smartphone
  • Reproducibility of Results
  • Ophthalmology & Optometry
  • Microscopy
  • Male
  • Humans
  • Healthy Volunteers
  • Female
  • Endothelium, Corneal
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Fliotsos, M. J., Deljookorani, S., Dzhaber, D., Chandan, S., Ighani, M., & Eghrari, A. O. (2020). Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy. Cornea, 39(7), 924–929. https://doi.org/10.1097/ico.0000000000002277
Fliotsos, Michael J., Shiva Deljookorani, Daliya Dzhaber, Subhangi Chandan, Mehrnaz Ighani, and Allen O. Eghrari. “Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy.Cornea 39, no. 7 (July 2020): 924–29. https://doi.org/10.1097/ico.0000000000002277.
Fliotsos MJ, Deljookorani S, Dzhaber D, Chandan S, Ighani M, Eghrari AO. Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy. Cornea. 2020 Jul;39(7):924–9.
Fliotsos, Michael J., et al. “Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy.Cornea, vol. 39, no. 7, July 2020, pp. 924–29. Epmc, doi:10.1097/ico.0000000000002277.
Fliotsos MJ, Deljookorani S, Dzhaber D, Chandan S, Ighani M, Eghrari AO. Qualitative and Quantitative Analysis of the Corneal Endothelium With Smartphone Specular Microscopy. Cornea. 2020 Jul;39(7):924–929.

Published In

Cornea

DOI

EISSN

1536-4798

ISSN

0277-3740

Publication Date

July 2020

Volume

39

Issue

7

Start / End Page

924 / 929

Related Subject Headings

  • Young Adult
  • Smartphone
  • Reproducibility of Results
  • Ophthalmology & Optometry
  • Microscopy
  • Male
  • Humans
  • Healthy Volunteers
  • Female
  • Endothelium, Corneal