Image restoration and spatial resolution in 7-tesla magnetic resonance imaging.

Publication ,  Journal Article
Lohmann, G; Bohn, S; Müller, K; Trampel, R; Turner, R
Published in: Magn Reson Med
July 2010

A good spatial resolution is essential for high precision segmentations of small structures in magnetic resonance images. However, any increase in the spatial resolution results in a decrease of the signal-to-noise ratio (SNR). In this article, this problem is addressed by a new image restoration technique that is used to partly compensate for the loss in SNR. Specifically, a two-stage hybrid image restoration procedure is proposed where the first stage is a Wiener wavelet filter for an initial denoising. The artifacts that will inevitably be produced by this step are subsequently reduced using a recent variant of anisotropic diffusion. The method is applied to magnetic resonance imaging data acquired on a 7-T magnetic resonance imaging scanner and compared with averaged multiple measurements of the same subject. It was found that the effect of image restoration procedure roughly corresponds to averaging across three repeated measurements.

Duke Scholars

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

July 2010

Volume

64

Issue

1

Start / End Page

15 / 22

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • Male
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Image Enhancement
  • Humans
  • Artifacts
  • Algorithms
  • Adult
  • 4003 Biomedical engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Lohmann, G., Bohn, S., Müller, K., Trampel, R., & Turner, R. (2010). Image restoration and spatial resolution in 7-tesla magnetic resonance imaging. Magn Reson Med, 64(1), 15–22. https://doi.org/10.1002/mrm.22488
Lohmann, Gabriele, Stefan Bohn, Karsten Müller, Robert Trampel, and Robert Turner. “Image restoration and spatial resolution in 7-tesla magnetic resonance imaging.Magn Reson Med 64, no. 1 (July 2010): 15–22. https://doi.org/10.1002/mrm.22488.
Lohmann G, Bohn S, Müller K, Trampel R, Turner R. Image restoration and spatial resolution in 7-tesla magnetic resonance imaging. Magn Reson Med. 2010 Jul;64(1):15–22.
Lohmann, Gabriele, et al. “Image restoration and spatial resolution in 7-tesla magnetic resonance imaging.Magn Reson Med, vol. 64, no. 1, July 2010, pp. 15–22. Pubmed, doi:10.1002/mrm.22488.
Lohmann G, Bohn S, Müller K, Trampel R, Turner R. Image restoration and spatial resolution in 7-tesla magnetic resonance imaging. Magn Reson Med. 2010 Jul;64(1):15–22.
Journal cover image

Published In

Magn Reson Med

DOI

EISSN

1522-2594

Publication Date

July 2010

Volume

64

Issue

1

Start / End Page

15 / 22

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • Male
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Image Enhancement
  • Humans
  • Artifacts
  • Algorithms
  • Adult
  • 4003 Biomedical engineering