A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States.
Journal articles
- Journal Article
Schultz, MK; Cessna, JT; Anderson, TL; Ponto, JA; Petry, N; Kowalsky, RJ; Palmer, MR; Beinlich, UF; Baker, W; Hinkle, GH; Hung, JC; Quinton, T ...
Published in: Appl Radiat Isot
February 2008
A blind performance test was conducted to evaluate dose-calibrator measurements at nuclear pharmacies in the United States (US). Two test-sample geometries were chosen to represent those used for measurements of 90Y-ibritumomab tiuxetan (ZEVALIN). The radioactivity concentration of test-samples was verified by the US National Institute of Standards and Technology. Forty-five results were reported by 10 participants. Eighty percent of reported values were within the US Pharmacopoeia content standard (+/-10%) for 90Y-ZEVALIN. All results were within US Nuclear Regulatory Commission conformance limits (+/-20%) for defining therapeutic misadministrations.
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
Appl Radiat Isot
DOI
ISSN
0969-8043
Publication Date
February 2008
Volume
66
Issue
2
Start / End Page
252 / 260
Location
England
Related Subject Headings
- Yttrium Radioisotopes
- United States
- Reference Standards
- Radiotherapy Dosage
- Radiopharmaceuticals
- Radiometry
- Radioimmunotherapy
- Quality Control
- Pilot Projects
- Pharmacies
Citation
APA
Chicago
ICMJE
MLA
NLM
Schultz, M. K., Cessna, J. T., Anderson, T. L., Ponto, J. A., Petry, N., Kowalsky, R. J., … Norenberg, J. P. (2008). A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States. Appl Radiat Isot, 66(2), 252–260. https://doi.org/10.1016/j.apradiso.2007.09.002
Schultz, Michael K., Jeffrey T. Cessna, Tamara L. Anderson, James A. Ponto, Neil Petry, Richard J. Kowalsky, Matthew R. Palmer, et al. “A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States.” Appl Radiat Isot 66, no. 2 (February 2008): 252–60. https://doi.org/10.1016/j.apradiso.2007.09.002.
Schultz MK, Cessna JT, Anderson TL, Ponto JA, Petry N, Kowalsky RJ, et al. A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States. Appl Radiat Isot. 2008 Feb;66(2):252–60.
Schultz, Michael K., et al. “A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States.” Appl Radiat Isot, vol. 66, no. 2, Feb. 2008, pp. 252–60. Pubmed, doi:10.1016/j.apradiso.2007.09.002.
Schultz MK, Cessna JT, Anderson TL, Ponto JA, Petry N, Kowalsky RJ, Palmer MR, Beinlich UF, Baker W, Hinkle GH, Hung JC, Quinton T, Rice PA, Divgi C, Norenberg JP. A performance evaluation of 90Y dose-calibrator measurements in nuclear pharmacies and clinics in the United States. Appl Radiat Isot. 2008 Feb;66(2):252–260.
Published In
Appl Radiat Isot
DOI
ISSN
0969-8043
Publication Date
February 2008
Volume
66
Issue
2
Start / End Page
252 / 260
Location
England
Related Subject Headings
- Yttrium Radioisotopes
- United States
- Reference Standards
- Radiotherapy Dosage
- Radiopharmaceuticals
- Radiometry
- Radioimmunotherapy
- Quality Control
- Pilot Projects
- Pharmacies