Validation of Macular Choroidal Thickness Measurements from Automated SD-OCT Image Segmentation.

Published

Journal Article

Spectral domain optical coherence tomography (SD-OCT) imaging permits in vivo visualization of the choroid with micron-level resolution over wide areas and is of interest for studies of ocular growth and myopia control. We evaluated the speed, repeatability, and accuracy of a new image segmentation method to quantify choroid thickness compared to manual segmentation.Two macular volumetric scans (25 × 30°) were taken from 30 eyes of 30 young adult subjects in two sessions, 1 hour apart. A single rater manually delineated choroid thickness as the distance between Bruch's membrane and sclera across three B-scans (foveal, inferior, and superior-most scan locations). Manual segmentation was compared to an automated method based on graph theory, dynamic programming, and wavelet-based texture analysis. Segmentation performance comparisons included processing speed, choroid thickness measurements across the foveal horizontal midline, and measurement repeatability (95% limits of agreement (LoA)).Subjects were healthy young adults (n = 30; 24 ± 2 years; mean ± SD; 63% female) with spherical equivalent refractive error of -3.46 ± 2.69D (range: +2.62 to -8.50D). Manual segmentation took 200 times longer than automated segmentation (780 vs. 4 seconds). Mean choroid thickness at the foveal center was 263 ± 24 μm (manual) and 259 ± 23 μm (automated), and this difference was not significant (p = 0.10). Regional segmentation errors across the foveal horizontal midline (±15°) were ≤9 μm (median) except for nasal-most regions closest to the nasal peripapillary margin-15 degrees (19 μm) and 12 degrees (16 μm) from the foveal center. Repeatability of choroidal thickness measurements had similar repeatability between segmentation methods (manual LoA: ±15 μm; automated LoA: ±14 μm).Automated segmentation of SD-OCT data by graph theory and dynamic programming is a fast, accurate, and reliable method to delineate the choroid. This approach will facilitate longitudinal studies evaluating changes in choroid thickness in response to novel optical corrections and in ocular disease.

Full Text

Duke Authors

Cited Authors

  • Twa, MD; Schulle, KL; Chiu, SJ; Farsiu, S; Berntsen, DA

Published Date

  • November 2016

Published In

Volume / Issue

  • 93 / 11

Start / End Page

  • 1387 - 1398

PubMed ID

  • 27668634

Pubmed Central ID

  • 27668634

Electronic International Standard Serial Number (EISSN)

  • 1538-9235

International Standard Serial Number (ISSN)

  • 1040-5488

Digital Object Identifier (DOI)

  • 10.1097/OPX.0000000000000985

Language

  • eng