Skip to main content

Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation.

Publication ,  Journal Article
Perez, MV; Mahaffey, KW; Hedlin, H; Rumsfeld, JS; Garcia, A; Ferris, T; Balasubramanian, V; Russo, AM; Rajmane, A; Cheung, L; Hung, G; Lee, J ...
Published in: N Engl J Med
November 14, 2019

BACKGROUND: Optical sensors on wearable devices can detect irregular pulses. The ability of a smartwatch application (app) to identify atrial fibrillation during typical use is unknown. METHODS: Participants without atrial fibrillation (as reported by the participants themselves) used a smartphone (Apple iPhone) app to consent to monitoring. If a smartwatch-based irregular pulse notification algorithm identified possible atrial fibrillation, a telemedicine visit was initiated and an electrocardiography (ECG) patch was mailed to the participant, to be worn for up to 7 days. Surveys were administered 90 days after notification of the irregular pulse and at the end of the study. The main objectives were to estimate the proportion of notified participants with atrial fibrillation shown on an ECG patch and the positive predictive value of irregular pulse intervals with a targeted confidence interval width of 0.10. RESULTS: We recruited 419,297 participants over 8 months. Over a median of 117 days of monitoring, 2161 participants (0.52%) received notifications of irregular pulse. Among the 450 participants who returned ECG patches containing data that could be analyzed - which had been applied, on average, 13 days after notification - atrial fibrillation was present in 34% (97.5% confidence interval [CI], 29 to 39) overall and in 35% (97.5% CI, 27 to 43) of participants 65 years of age or older. Among participants who were notified of an irregular pulse, the positive predictive value was 0.84 (95% CI, 0.76 to 0.92) for observing atrial fibrillation on the ECG simultaneously with a subsequent irregular pulse notification and 0.71 (97.5% CI, 0.69 to 0.74) for observing atrial fibrillation on the ECG simultaneously with a subsequent irregular tachogram. Of 1376 notified participants who returned a 90-day survey, 57% contacted health care providers outside the study. There were no reports of serious app-related adverse events. CONCLUSIONS: The probability of receiving an irregular pulse notification was low. Among participants who received notification of an irregular pulse, 34% had atrial fibrillation on subsequent ECG patch readings and 84% of notifications were concordant with atrial fibrillation. This siteless (no on-site visits were required for the participants), pragmatic study design provides a foundation for large-scale pragmatic studies in which outcomes or adherence can be reliably assessed with user-owned devices. (Funded by Apple; Apple Heart Study ClinicalTrials.gov number, NCT03335800.).

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

N Engl J Med

DOI

EISSN

1533-4406

Publication Date

November 14, 2019

Volume

381

Issue

20

Start / End Page

1909 / 1917

Location

United States

Related Subject Headings

  • Wearable Electronic Devices
  • Telemedicine
  • Prospective Studies
  • Predictive Value of Tests
  • Mobile Applications
  • Middle Aged
  • Male
  • Humans
  • General & Internal Medicine
  • Female
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Perez, M. V., Mahaffey, K. W., Hedlin, H., Rumsfeld, J. S., Garcia, A., Ferris, T., … Apple Heart Study Investigators, . (2019). Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med, 381(20), 1909–1917. https://doi.org/10.1056/NEJMoa1901183
Perez, Marco V., Kenneth W. Mahaffey, Haley Hedlin, John S. Rumsfeld, Ariadna Garcia, Todd Ferris, Vidhya Balasubramanian, et al. “Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation.N Engl J Med 381, no. 20 (November 14, 2019): 1909–17. https://doi.org/10.1056/NEJMoa1901183.
Perez MV, Mahaffey KW, Hedlin H, Rumsfeld JS, Garcia A, Ferris T, et al. Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019 Nov 14;381(20):1909–17.
Perez, Marco V., et al. “Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation.N Engl J Med, vol. 381, no. 20, Nov. 2019, pp. 1909–17. Pubmed, doi:10.1056/NEJMoa1901183.
Perez MV, Mahaffey KW, Hedlin H, Rumsfeld JS, Garcia A, Ferris T, Balasubramanian V, Russo AM, Rajmane A, Cheung L, Hung G, Lee J, Kowey P, Talati N, Nag D, Gummidipundi SE, Beatty A, Hills MT, Desai S, Granger CB, Desai M, Turakhia MP, Apple Heart Study Investigators. Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019 Nov 14;381(20):1909–1917.

Published In

N Engl J Med

DOI

EISSN

1533-4406

Publication Date

November 14, 2019

Volume

381

Issue

20

Start / End Page

1909 / 1917

Location

United States

Related Subject Headings

  • Wearable Electronic Devices
  • Telemedicine
  • Prospective Studies
  • Predictive Value of Tests
  • Mobile Applications
  • Middle Aged
  • Male
  • Humans
  • General & Internal Medicine
  • Female