Performance standards in positron emission tomography.

Published

Journal Article

A standard set of performance measurements is proposed for use with positron emission tomographs. This set of measurements has been developed jointly by the Computer and Instrumentation Council of the Society of Nuclear Medicine and the National Electrical Manufacturers Association. The measurements include tests of spatial resolution, scatter fraction, sensitivity, count rate losses and randoms, uniformity, scatter correction, attenuation correction, and count rate linearity correction.

Full Text

Duke Authors

Cited Authors

  • Karp, JS; Daube-Witherspoon, ME; Hoffman, EJ; Lewellen, TK; Links, JM; Wong, WH; Hichwa, RD; Casey, ME; Colsher, JG; Hitchens, RE

Published Date

  • December 1991

Published In

Volume / Issue

  • 32 / 12

Start / End Page

  • 2342 - 2350

PubMed ID

  • 1744726

Pubmed Central ID

  • 1744726

International Standard Serial Number (ISSN)

  • 0161-5505

Language

  • eng

Conference Location

  • United States