Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation.

Publication ,  Journal Article
Howell, B; Naik, S; Grill, WM
Published in: IEEE transactions on bio-medical engineering
February 2014

Deep brain stimulation (DBS) is an established therapy for movement disorders, but the fundamental mechanisms by which DBS has its effects remain unknown. Computational models can provide insights into the mechanisms of DBS, but to be useful, the models must have sufficient detail to predict accurately the electric fields produced by DBS. We used a finite-element method model of the Medtronic 3387 electrode array, coupled to cable models of myelinated axons, to quantify how interpolation errors, electrode geometry, and the electrode-tissue interface affect calculation of electrical potentials and stimulation thresholds for populations of model nerve fibers. Convergence of the potentials was not a sufficient criterion for ensuring the same degree of accuracy in subsequent determination of stimulation thresholds, because the accuracy of the stimulation thresholds depended on the order of the elements. Simplifying the 3387 electrode array by ignoring the inactive contacts and extending the terminated end of the shaft had position-dependent effects on the potentials and excitation thresholds, and these simplifications may impact correlations between DBS parameters and clinical outcomes. When the current density in the bulk tissue is uniform, the effect of the electrode-tissue interface impedance could be approximated by filtering the potentials calculated with a static lumped electrical equivalent circuit. Further, for typical DBS parameters during voltage-regulated stimulation, it was valid to approximate the electrode as an ideal polarized electrode with a nonlinear capacitance. Validation of these computational considerations enables accurate modeling of the electric field produced by DBS.

Duke Scholars

Published In

IEEE transactions on bio-medical engineering

DOI

EISSN

1558-2531

ISSN

0018-9294

Publication Date

February 2014

Volume

61

Issue

2

Start / End Page

297 / 307

Related Subject Headings

  • Models, Neurological
  • Humans
  • Finite Element Analysis
  • Electrodes
  • Deep Brain Stimulation
  • Biomedical Engineering
  • Biomedical Engineering
  • 4603 Computer vision and multimedia computation
  • 4009 Electronics, sensors and digital hardware
  • 4003 Biomedical engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Howell, B., Naik, S., & Grill, W. M. (2014). Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation. IEEE Transactions on Bio-Medical Engineering, 61(2), 297–307. https://doi.org/10.1109/tbme.2013.2292025
Howell, Bryan, Sagar Naik, and Warren M. Grill. “Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation.IEEE Transactions on Bio-Medical Engineering 61, no. 2 (February 2014): 297–307. https://doi.org/10.1109/tbme.2013.2292025.
Howell, Bryan, et al. “Influences of interpolation error, electrode geometry, and the electrode-tissue interface on models of electric fields produced by deep brain stimulation.IEEE Transactions on Bio-Medical Engineering, vol. 61, no. 2, Feb. 2014, pp. 297–307. Epmc, doi:10.1109/tbme.2013.2292025.

Published In

IEEE transactions on bio-medical engineering

DOI

EISSN

1558-2531

ISSN

0018-9294

Publication Date

February 2014

Volume

61

Issue

2

Start / End Page

297 / 307

Related Subject Headings

  • Models, Neurological
  • Humans
  • Finite Element Analysis
  • Electrodes
  • Deep Brain Stimulation
  • Biomedical Engineering
  • Biomedical Engineering
  • 4603 Computer vision and multimedia computation
  • 4009 Electronics, sensors and digital hardware
  • 4003 Biomedical engineering