RATIONALE: Hyperpolarized 129Xe MRI enables quantitative evaluation of regional ventilation. To this end, multiple classifiers have been proposed to determine ventilation defect percentage (VDP) as well as other cluster populations. However, consensus has not yet been reached regarding which of these methods to deploy for multicenter clinical trials. Here, we compare two published classification techniques-linear-binning and adaptive K-means-to establish their limits of agreement and their robustness against reduced signal-to-noise ratio (SNR). METHODS: A total of 29 subjects (age: 38.4 ± 19.0 years) were retrospectively identified for inter-method comparison. For each 129Xe ventilation image, 7 images with reduced SNR were generated with equal decrements relative to the native SNR. All 8 sets of images were then analyzed using both methods independently to classify all lung voxels into four clusters: VDP, low-, medium-, and high-ventilation-percentage (LVP, MVP and HVP). For each cluster, the percentage of the lung it comprised was compared between the two methods, as well as how these values persisted as SNR was degraded. RESULTS: The limits of agreement for calculating VDP were [+0.2%, +4.0%] with a +1.5% bias for binning relative to K-means. However, the inter-method agreement for the other clusters was moderate, with biases of -5.7%, 8.1%, and -4.0% for LVP, MVP, and HVP, respectively. As SNR decreased below ∼4, both methods began reporting values that deviated substantially from the native image. By requiring VDP to remain within ≤1.8% of that calculated from the native image, the minimum tolerable SNR values were 2.4 ± 1.0 for the linear-binning, and 3.5 ± 1.5 for the K-means. CONCLUSIONS: Both methods agree well in quantifying VDP, but agreement for LVP and MVP remains variable. We suggest a required SNR threshold be two standard deviations above the minimum value of 3.5 ± 1.5 for robust determination of VDP, suggesting a minimum SNR of 6.6. However, robust quantification of the ventilated clusters required an SNR of 13.4.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Acad Radiol

DOI

EISSN

1878-4046

Publication Date

July 2019

Volume

26

Issue

7

Start / End Page

949 / 959

Location

United States

Related Subject Headings

  • Young Adult
  • Xenon Isotopes
  • Signal-To-Noise Ratio
  • Retrospective Studies
  • Pulmonary Ventilation
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Magnetic Resonance Imaging
  • Lung
 

Citation

APA
Chicago
ICMJE
MLA
NLM
He, M., Zha, W., Tan, F., Rankine, L., Fain, S., & Driehuys, B. (2019). A Comparison of Two Hyperpolarized 129Xe MRI Ventilation Quantification Pipelines: The Effect of Signal to Noise Ratio. Acad Radiol, 26(7), 949–959. https://doi.org/10.1016/j.acra.2018.08.015
He, Mu, Wei Zha, Fei Tan, Leith Rankine, Sean Fain, and Bastiaan Driehuys. “A Comparison of Two Hyperpolarized 129Xe MRI Ventilation Quantification Pipelines: The Effect of Signal to Noise Ratio.Acad Radiol 26, no. 7 (July 2019): 949–59. https://doi.org/10.1016/j.acra.2018.08.015.
He M, Zha W, Tan F, Rankine L, Fain S, Driehuys B. A Comparison of Two Hyperpolarized 129Xe MRI Ventilation Quantification Pipelines: The Effect of Signal to Noise Ratio. Acad Radiol. 2019 Jul;26(7):949–59.
He, Mu, et al. “A Comparison of Two Hyperpolarized 129Xe MRI Ventilation Quantification Pipelines: The Effect of Signal to Noise Ratio.Acad Radiol, vol. 26, no. 7, July 2019, pp. 949–59. Pubmed, doi:10.1016/j.acra.2018.08.015.
He M, Zha W, Tan F, Rankine L, Fain S, Driehuys B. A Comparison of Two Hyperpolarized 129Xe MRI Ventilation Quantification Pipelines: The Effect of Signal to Noise Ratio. Acad Radiol. 2019 Jul;26(7):949–959.
Journal cover image

Published In

Acad Radiol

DOI

EISSN

1878-4046

Publication Date

July 2019

Volume

26

Issue

7

Start / End Page

949 / 959

Location

United States

Related Subject Headings

  • Young Adult
  • Xenon Isotopes
  • Signal-To-Noise Ratio
  • Retrospective Studies
  • Pulmonary Ventilation
  • Nuclear Medicine & Medical Imaging
  • Middle Aged
  • Male
  • Magnetic Resonance Imaging
  • Lung