Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method.

Publication ,  Journal Article
Menaesse, A; Sumetsky, D; Emanuely, N; Stein, JL; Gates, EM; Hoffman, BD; Boustany, NN
Published in: Cytometry. Part A : the journal of the International Society for Analytical Cytology
April 2021

Fӧrster (or fluorescence) resonance energy transfer (FRET) is a quantifiable energy transfer in which a donor fluorophore nonradiatively transfers its excitation energy to an acceptor fluorophore. A change in FRET efficiency indicates a change of proximity and environment of these fluorophores, which enables the study of intermolecular interactions. Measurement of FRET efficiency using the sensitized emission method requires a donor-acceptor calibrated system. One of these calibration factors named the G factor, which depends on instrument parameters related to the donor and acceptor measurement channels and on the fluorophores quantum efficiencies, can be determined in several different ways and allows for conversion of the raw donor and acceptor emission signals to FRET efficiency. However, the calculated value of the G factor from experimental data can fluctuate significantly depending on the chosen experimental method and the size of the sample. In this technical note, we extend the results of Gates et al. (Cytometry Part A 95A (2018) 201-213) by refining the calibration method used for calibration of FRET from image pixel data. Instead of using the pixel histograms of two constructs with high and low FRET efficiency to determine the G factor, we use pixel histogram data from one construct of known efficiency. We validate this method by determining the G factor with the same constructs developed and used by Gates et al. and comparing the results from the two approaches. While the two approaches are equivalent theoretically, we demonstrate that the use of a single construct with known efficiency provides a more precise experimental measurement of the G factor that can be attained by collecting a smaller number of images. © 2020 International Society for Advancement of Cytometry.

Duke Scholars

Published In

Cytometry. Part A : the journal of the International Society for Analytical Cytology

DOI

EISSN

1552-4930

ISSN

1552-4922

Publication Date

April 2021

Volume

99

Issue

4

Start / End Page

407 / 416

Related Subject Headings

  • Immunology
  • Fluorescent Dyes
  • Fluorescence Resonance Energy Transfer
  • Calibration
  • 3201 Cardiovascular medicine and haematology
  • 0601 Biochemistry and Cell Biology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Menaesse, A., Sumetsky, D., Emanuely, N., Stein, J. L., Gates, E. M., Hoffman, B. D., & Boustany, N. N. (2021). Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method. Cytometry. Part A : The Journal of the International Society for Analytical Cytology, 99(4), 407–416. https://doi.org/10.1002/cyto.a.24194
Menaesse, Audrey, Daniel Sumetsky, Nicolas Emanuely, Jeremy L. Stein, Evan M. Gates, Brenton D. Hoffman, and Nada N. Boustany. “Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method.Cytometry. Part A : The Journal of the International Society for Analytical Cytology 99, no. 4 (April 2021): 407–16. https://doi.org/10.1002/cyto.a.24194.
Menaesse A, Sumetsky D, Emanuely N, Stein JL, Gates EM, Hoffman BD, et al. Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method. Cytometry Part A : the journal of the International Society for Analytical Cytology. 2021 Apr;99(4):407–16.
Menaesse, Audrey, et al. “Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method.Cytometry. Part A : The Journal of the International Society for Analytical Cytology, vol. 99, no. 4, Apr. 2021, pp. 407–16. Epmc, doi:10.1002/cyto.a.24194.
Menaesse A, Sumetsky D, Emanuely N, Stein JL, Gates EM, Hoffman BD, Boustany NN. Simplified Instrument Calibration for Wide-Field Fluorescence Resonance Energy Transfer (FRET) Measured by the Sensitized Emission Method. Cytometry Part A : the journal of the International Society for Analytical Cytology. 2021 Apr;99(4):407–416.
Journal cover image

Published In

Cytometry. Part A : the journal of the International Society for Analytical Cytology

DOI

EISSN

1552-4930

ISSN

1552-4922

Publication Date

April 2021

Volume

99

Issue

4

Start / End Page

407 / 416

Related Subject Headings

  • Immunology
  • Fluorescent Dyes
  • Fluorescence Resonance Energy Transfer
  • Calibration
  • 3201 Cardiovascular medicine and haematology
  • 0601 Biochemistry and Cell Biology