Skip to main content

High-Resolution Power Doppler Using Null Subtraction Imaging.

Publication ,  Journal Article
Kou, Z; Lowerison, MR; You, Q; Wang, Y; Song, P; Oelze, ML
Published in: IEEE transactions on medical imaging
September 2024

To improve the spatial resolution of power Doppler (PD) imaging, we explored null subtraction imaging (NSI) as an alternative beamforming technique to delay-and-sum (DAS). NSI is a nonlinear beamforming approach that uses three different apodizations on receive and incoherently sums the beamformed envelopes. NSI uses a null in the beam pattern to improve the lateral resolution, which we apply here for improving PD spatial resolution both with and without contrast microbubbles. In this study, we used NSI with three types of singular value decomposition (SVD)-based clutter filters and noise equalization to generate high-resolution PD images. An element sensitivity correction scheme was also proposed as a crucial component of NSI-based PD imaging. First, a microbubble trace experiment was performed to evaluate the resolution improvement of NSI-based PD over traditional DAS-based PD. Then, both contrast-enhanced and contrast free ultrasound PD images were generated from the scan of a rat brain. The cross-sectional profile of the microbubble traces and microvessels were plotted. FWHM was also estimated to provide a quantitative metric. Furthermore, iso-frequency curves were calculated to provide a resolution evaluation metric over the global field of view. Up to six-fold resolution improvement was demonstrated by the FWHM estimate and four-fold resolution improvement was demonstrated by the iso-frequency curve from the NSI-based PD microvessel images compared to microvessel images generated by traditional DAS-based beamforming. A resolvability of [Formula: see text] was measured from the NSI-based PD microvessel image. The computational cost of NSI-based PD was only increased by 40 percent over the DAS-based PD.

Duke Scholars

Published In

IEEE transactions on medical imaging

DOI

EISSN

1558-254X

ISSN

0278-0062

Publication Date

September 2024

Volume

43

Issue

9

Start / End Page

3060 / 3071

Related Subject Headings

  • Ultrasonography, Doppler
  • Subtraction Technique
  • Signal Processing, Computer-Assisted
  • Rats, Sprague-Dawley
  • Rats
  • Nuclear Medicine & Medical Imaging
  • Microbubbles
  • Image Processing, Computer-Assisted
  • Contrast Media
  • Brain
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Kou, Z., Lowerison, M. R., You, Q., Wang, Y., Song, P., & Oelze, M. L. (2024). High-Resolution Power Doppler Using Null Subtraction Imaging. IEEE Transactions on Medical Imaging, 43(9), 3060–3071. https://doi.org/10.1109/tmi.2024.3383768
Kou, Zhengchang, Matthew R. Lowerison, Qi You, Yike Wang, Pengfei Song, and Michael L. Oelze. “High-Resolution Power Doppler Using Null Subtraction Imaging.IEEE Transactions on Medical Imaging 43, no. 9 (September 2024): 3060–71. https://doi.org/10.1109/tmi.2024.3383768.
Kou Z, Lowerison MR, You Q, Wang Y, Song P, Oelze ML. High-Resolution Power Doppler Using Null Subtraction Imaging. IEEE transactions on medical imaging. 2024 Sep;43(9):3060–71.
Kou, Zhengchang, et al. “High-Resolution Power Doppler Using Null Subtraction Imaging.IEEE Transactions on Medical Imaging, vol. 43, no. 9, Sept. 2024, pp. 3060–71. Epmc, doi:10.1109/tmi.2024.3383768.
Kou Z, Lowerison MR, You Q, Wang Y, Song P, Oelze ML. High-Resolution Power Doppler Using Null Subtraction Imaging. IEEE transactions on medical imaging. 2024 Sep;43(9):3060–3071.

Published In

IEEE transactions on medical imaging

DOI

EISSN

1558-254X

ISSN

0278-0062

Publication Date

September 2024

Volume

43

Issue

9

Start / End Page

3060 / 3071

Related Subject Headings

  • Ultrasonography, Doppler
  • Subtraction Technique
  • Signal Processing, Computer-Assisted
  • Rats, Sprague-Dawley
  • Rats
  • Nuclear Medicine & Medical Imaging
  • Microbubbles
  • Image Processing, Computer-Assisted
  • Contrast Media
  • Brain