Electrical properties of implant encapsulation tissue.

Publication ,  Journal Article
Grill, WM; Mortimer, JT
Published in: Annals of biomedical engineering
January 1994

The purpose of this study was to determine the electrical properties of the encapsulation tissue that surrounds electrodes chronically implanted in the body. Two four-electrode arrays, fabricated from either epoxy or silicone rubber, were implanted in each of six adult cats for 82 to 156 days. In vivo measurements of tissue resistivity using the four-electrode technique indicated that formation of the encapsulation tissue resulted in a significant increase in the resistivity of the tissue around the arrays. In vitro measurements of tissue impedance using a four-electrode cell indicated that the resistivity of the encapsulation tissue was a function of the tissue morphology. The tight layers of fibroblasts and collagen that formed around the silicone rubber arrays had a resistivity of 627 +/- 108 omega-cm (mean +/- SD; n = 6), which was independent of frequency from 10 Hz to 100 kHz, and was significantly larger than the resistivity of the epoxy encapsulation tissue at all frequencies between 20 Hz and 100 kHz. The combination of macrophages, foreign body giant cells, loose collagen, and fibroblasts that formed around the epoxy arrays had a frequency-dependent resistivity that decreased from 454 +/- 123 omega-cm (n = 5) to 193 +/- 98 omega-cm between 10 Hz and 1 kHz, and was independent of frequency between 1 kHz and 100 kHz, with a mean value of 195 +/- 88 omega-cm. The results indicate that the resistivity of the encapsulation tissue is sufficient to alter the shape and magnitude of the electric field generated by chronically implanted electrodes.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Annals of biomedical engineering

DOI

EISSN

1573-9686

ISSN

0090-6964

Publication Date

January 1994

Volume

22

Issue

1

Start / End Page

23 / 33

Related Subject Headings

  • Macrophages
  • Foreign-Body Reaction
  • Fibroblasts
  • Evaluation Studies as Topic
  • Electrophysiology
  • Electrodes, Implanted
  • Electric Impedance
  • Electric Conductivity
  • Collagen
  • Chronic Disease
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Grill, W. M., & Mortimer, J. T. (1994). Electrical properties of implant encapsulation tissue. Annals of Biomedical Engineering, 22(1), 23–33. https://doi.org/10.1007/bf02368219
Grill, W. M., and J. T. Mortimer. “Electrical properties of implant encapsulation tissue.Annals of Biomedical Engineering 22, no. 1 (January 1994): 23–33. https://doi.org/10.1007/bf02368219.
Grill WM, Mortimer JT. Electrical properties of implant encapsulation tissue. Annals of biomedical engineering. 1994 Jan;22(1):23–33.
Grill, W. M., and J. T. Mortimer. “Electrical properties of implant encapsulation tissue.Annals of Biomedical Engineering, vol. 22, no. 1, Jan. 1994, pp. 23–33. Epmc, doi:10.1007/bf02368219.
Grill WM, Mortimer JT. Electrical properties of implant encapsulation tissue. Annals of biomedical engineering. 1994 Jan;22(1):23–33.
Journal cover image

Published In

Annals of biomedical engineering

DOI

EISSN

1573-9686

ISSN

0090-6964

Publication Date

January 1994

Volume

22

Issue

1

Start / End Page

23 / 33

Related Subject Headings

  • Macrophages
  • Foreign-Body Reaction
  • Fibroblasts
  • Evaluation Studies as Topic
  • Electrophysiology
  • Electrodes, Implanted
  • Electric Impedance
  • Electric Conductivity
  • Collagen
  • Chronic Disease