Skip to main content

Lead design and initial applications of a new lead for long-term endovascular vagal stimulation.

Publication ,  Journal Article
Nabutovsky, Y; Florio, J; Morgan, K; Grill, WM; Farazi, TG
Published in: Pacing and clinical electrophysiology : PACE
January 2007

Vagal nerve stimulation (VNS) has negative chronotropic and dromotropic effects. We developed and tested an endovascular spiral vagal stimulation lead (ESVL) designed to follow the projection of the cardiac branches of the vagus nerve around the superior vena cava (SVC) to optimize VNS.ESVL contained six 5-mm coil electrodes, spaced 5-mm apart with a spiral guidewire to provide shape. The tightness and diameter of the guidewire were changed before each placement to simulate different lead designs. Various 2-, 3-, and 4-electrode combinations were used and several lead positions were tested each time. Each VNS protocol included 2-12 V, 15-second pulse trains at 20 Hz, with 2 ms pulse duration. A basket catheter (BC) was used as control and to approximate the initial VNS location. The VNS protocol was performed at the optimal location, using first the BC and then several ESVL configurations.VNS caused a voltage-dependent decrease in heart rate (HR). Using the optimal ESVL configuration at 7 V, HR decreased by 30.4% (37.2 bpm) in dog no. 1 and 12.4% (16.6 bpm) in dog no. 2, versus 15.5% (16.6 bpm) and 16.7% (19.5 bpm) with the BC.A new endovascular spiral lead that takes advantage of the anatomy of the cardiac branches of the vagus nerve in the SVC was developed. VNS using ESVL produced significant HR slowing at voltages slightly below the highest pulse generator output of 7.5 V, which may be suitable for long-term implantation.

Duke Scholars

Published In

Pacing and clinical electrophysiology : PACE

DOI

EISSN

1540-8159

ISSN

0147-8389

Publication Date

January 2007

Volume

30 Suppl 1

Start / End Page

S215 / S218

Related Subject Headings

  • Vena Cava, Superior
  • Vagus Nerve
  • Heart Rate
  • Equipment Design
  • Electrodes
  • Electric Stimulation Therapy
  • Dogs
  • Cardiovascular System & Hematology
  • Animals
  • 3202 Clinical sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Nabutovsky, Y., Florio, J., Morgan, K., Grill, W. M., & Farazi, T. G. (2007). Lead design and initial applications of a new lead for long-term endovascular vagal stimulation. Pacing and Clinical Electrophysiology : PACE, 30 Suppl 1, S215–S218. https://doi.org/10.1111/j.1540-8159.2007.00641.x
Nabutovsky, Yelena, Joseph Florio, Kevin Morgan, Warren M. Grill, and Taraneh G. Farazi. “Lead design and initial applications of a new lead for long-term endovascular vagal stimulation.Pacing and Clinical Electrophysiology : PACE 30 Suppl 1 (January 2007): S215–18. https://doi.org/10.1111/j.1540-8159.2007.00641.x.
Nabutovsky Y, Florio J, Morgan K, Grill WM, Farazi TG. Lead design and initial applications of a new lead for long-term endovascular vagal stimulation. Pacing and clinical electrophysiology : PACE. 2007 Jan;30 Suppl 1:S215–8.
Nabutovsky, Yelena, et al. “Lead design and initial applications of a new lead for long-term endovascular vagal stimulation.Pacing and Clinical Electrophysiology : PACE, vol. 30 Suppl 1, Jan. 2007, pp. S215–18. Epmc, doi:10.1111/j.1540-8159.2007.00641.x.
Nabutovsky Y, Florio J, Morgan K, Grill WM, Farazi TG. Lead design and initial applications of a new lead for long-term endovascular vagal stimulation. Pacing and clinical electrophysiology : PACE. 2007 Jan;30 Suppl 1:S215–S218.

Published In

Pacing and clinical electrophysiology : PACE

DOI

EISSN

1540-8159

ISSN

0147-8389

Publication Date

January 2007

Volume

30 Suppl 1

Start / End Page

S215 / S218

Related Subject Headings

  • Vena Cava, Superior
  • Vagus Nerve
  • Heart Rate
  • Equipment Design
  • Electrodes
  • Electric Stimulation Therapy
  • Dogs
  • Cardiovascular System & Hematology
  • Animals
  • 3202 Clinical sciences