Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation.

Publication ,  Journal Article
Patel, SG; Malkin, RA
Published in: IEEE Trans Biomed Eng
October 2007

Unintended, weak AC stimulation (leakage currents) from medical devices can cause blood pressure collapse and ventricular fibrillation (VF), potentially even death. Yet, little is understood about AC cardiac stimulation. The objective of this paper is to establish the relationship between the stimulation and VF thresholds for electrode size and stimulation frequency. Twenty-four retired male breeder guinea pigs were anesthetized with isoflurane, a tracheotomy and thoracotomy were performed, and vitals were monitored using the lead II ECG and an optical plethysmograph. The circular flat ends of eleven stainless steel rods were used as electrodes with areas ranging from 0.1 to 26.79 mm2. In the first study, 60-Hz AC stimuli of 5 s duration were delivered with strengths from 25-3000 microA or until VF was induced. In the second group, the current thresholds at 20, 40, 80, and 160 Hz were determined at electrode areas of 0.2, 2.01, and 16.4 mm2. Reactions were categorized as having no effect, having some effect (EFFECT, typically blood pressure collapse), and inducing VF. On a log-log scale, electrode radii had a piecewise-linear relationship with the current thresholds for EFFECT (p

Duke Scholars

Published In

IEEE Trans Biomed Eng

DOI

ISSN

0018-9294

Publication Date

October 2007

Volume

54

Issue

10

Start / End Page

1829 / 1836

Related Subject Headings

  • Ventricular Fibrillation
  • Risk Factors
  • Risk Assessment
  • Microelectrodes
  • Male
  • Guinea Pigs
  • Equipment Failure Analysis
  • Equipment Design
  • Electricity
  • Electric Stimulation
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Patel, S. G., & Malkin, R. A. (2007). Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation. IEEE Trans Biomed Eng, 54(10), 1829–1836. https://doi.org/10.1109/TBME.2007.904528
Patel, Shivani Govind, and Robert A. Malkin. “Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation.IEEE Trans Biomed Eng 54, no. 10 (October 2007): 1829–36. https://doi.org/10.1109/TBME.2007.904528.
Patel SG, Malkin RA. Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation. IEEE Trans Biomed Eng. 2007 Oct;54(10):1829–36.
Patel, Shivani Govind, and Robert A. Malkin. “Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation.IEEE Trans Biomed Eng, vol. 54, no. 10, Oct. 2007, pp. 1829–36. Manual, doi:10.1109/TBME.2007.904528.
Patel SG, Malkin RA. Effect of electrode surface area on thresholds for AC stimulation and ventricular fibrillation. IEEE Trans Biomed Eng. 2007 Oct;54(10):1829–1836.

Published In

IEEE Trans Biomed Eng

DOI

ISSN

0018-9294

Publication Date

October 2007

Volume

54

Issue

10

Start / End Page

1829 / 1836

Related Subject Headings

  • Ventricular Fibrillation
  • Risk Factors
  • Risk Assessment
  • Microelectrodes
  • Male
  • Guinea Pigs
  • Equipment Failure Analysis
  • Equipment Design
  • Electricity
  • Electric Stimulation