Skip to main content

Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety.

Publication ,  Journal Article
Swerdlow, CD; Olson, WH; O'Connor, ME; Gallik, DM; Malkin, RA; Laks, M
Published in: Circulation
May 1999

The national standard for safe 60-Hz intracardiac leakage current under a single-fault condition is 50 microA. This standard is intended to protect patients from alternating current (AC) at levels below the threshold for sensation, but the minimum unsafe level for AC in closed-chest humans is not known. To determine this value, we studied 40 patients at testing of implantable cardioverter-defibrillators using a programmable source of 60-Hz AC.We applied AC for 5-second test periods in increasing strengths until ventricular fibrillation (VF) was induced or 1 mA was reached. Two current paths were tested: bipolar, between tip and ring electrodes of a right ventricular pacing catheter, and unipolar, from tip to a remote electrode. We observed a characteristic sequence of 3 responses as AC was increased: (1) intermittent ventricular capture with QRS morphology identical to pacing through the electrodes (minimum value, 20 microA); (2) continuous capture at cycle length 282+/-88 ms (minimum value, 32 microA); and (3) VF persisting after AC termination (minimum value, 49 microA). Continuous capture caused loss of pulsatile arterial pressure and cardiovascular collapse (mean arterial pressure, 32+/-8 mm Hg) for the duration of AC with no ECG evidence of AC stimulation. Thus, the clinical picture was that of hypotensive ventricular tachycardia (VT). The continuous-capture threshold was ConclusionsLeakage current causes cardiovascular collapse at levels below the VF threshold. Stimulation by silent AC that is neither felt nor visible on the ECG presents as hypotensive VT. In patients with intracardiac electrodes, leakage current less than or equal to the present standard of 50 microA may cause VT or VF. The safety standard for leakage current lasting >/=5 seconds should be

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Circulation

DOI

EISSN

1524-4539

ISSN

0009-7322

Publication Date

May 1999

Volume

99

Issue

19

Start / End Page

2559 / 2564

Related Subject Headings

  • Ventricular Fibrillation
  • Humans
  • Heart
  • Electrocardiography
  • Electric Stimulation
  • Cardiovascular System & Hematology
  • 4207 Sports science and exercise
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology
  • 1117 Public Health and Health Services
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Swerdlow, C. D., Olson, W. H., O’Connor, M. E., Gallik, D. M., Malkin, R. A., & Laks, M. (1999). Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety. Circulation, 99(19), 2559–2564. https://doi.org/10.1161/01.cir.99.19.2559
Swerdlow, C. D., W. H. Olson, M. E. O’Connor, D. M. Gallik, R. A. Malkin, and M. Laks. “Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety.Circulation 99, no. 19 (May 1999): 2559–64. https://doi.org/10.1161/01.cir.99.19.2559.
Swerdlow CD, Olson WH, O’Connor ME, Gallik DM, Malkin RA, Laks M. Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety. Circulation. 1999 May;99(19):2559–64.
Swerdlow, C. D., et al. “Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety.Circulation, vol. 99, no. 19, May 1999, pp. 2559–64. Epmc, doi:10.1161/01.cir.99.19.2559.
Swerdlow CD, Olson WH, O’Connor ME, Gallik DM, Malkin RA, Laks M. Cardiovascular collapse caused by electrocardiographically silent 60-Hz intracardiac leakage current. Implications for electrical safety. Circulation. 1999 May;99(19):2559–2564.

Published In

Circulation

DOI

EISSN

1524-4539

ISSN

0009-7322

Publication Date

May 1999

Volume

99

Issue

19

Start / End Page

2559 / 2564

Related Subject Headings

  • Ventricular Fibrillation
  • Humans
  • Heart
  • Electrocardiography
  • Electric Stimulation
  • Cardiovascular System & Hematology
  • 4207 Sports science and exercise
  • 3202 Clinical sciences
  • 3201 Cardiovascular medicine and haematology
  • 1117 Public Health and Health Services