On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT.

Publication ,  Journal Article
Pierquet, M; Craciunescu, O; Steffey, B; Song, H; Oldham, M
Published in: J Phys Conf Ser
December 7, 2010

PURPOSE: The feasibility of using the PRESAGE/Optical-CT system for 3D dosimetry verification around a brachytherapy source is investigated. METHOD AND MATERIALS: Brachytherapy dose distributions were obtained by irradiation of cylindrical PRESAGE volumes 6cm in diameter by 8cm height with a GammaMed 12i Ir-192 HDR unit (Varian Medical Systems). A narrow channel on the central axis was created by setting a steel catheter in the Presage during manufacture, enabling measurements close to the source (~3mm). RESULTS: Comparison of dose line profiles shows good agreement between PRESAGE and verified calculated dose calculation, in both high and low dose regions. CONCLUSION: The PRESAGE/Optical-CT shows good potential in verification of 3D dose distributions around brachytherapy sources.

Duke Scholars

Published In

J Phys Conf Ser

DOI

ISSN

1742-6588

Publication Date

December 7, 2010

Volume

250

Issue

1

Start / End Page

120911 / 120915

Location

England

Related Subject Headings

  • 51 Physical sciences
  • 0299 Other Physical Sciences
  • 0204 Condensed Matter Physics
  • 0202 Atomic, Molecular, Nuclear, Particle and Plasma Physics
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Pierquet, M., Craciunescu, O., Steffey, B., Song, H., & Oldham, M. (2010). On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT. J Phys Conf Ser, 250(1), 120911–120915. https://doi.org/10.1088/1742-6596/250/1/012091
Pierquet, M., O. Craciunescu, B. Steffey, H. Song, and M. Oldham. “On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT.J Phys Conf Ser 250, no. 1 (December 7, 2010): 120911–15. https://doi.org/10.1088/1742-6596/250/1/012091.
Pierquet M, Craciunescu O, Steffey B, Song H, Oldham M. On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT. J Phys Conf Ser. 2010 Dec 7;250(1):120911–5.
Pierquet, M., et al. “On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT.J Phys Conf Ser, vol. 250, no. 1, Dec. 2010, pp. 120911–15. Pubmed, doi:10.1088/1742-6596/250/1/012091.
Pierquet M, Craciunescu O, Steffey B, Song H, Oldham M. On the Feasibility of Verification of 3D Dosimetry Near Brachytherapy Sources Using PRESAGE/Optical-CT. J Phys Conf Ser. 2010 Dec 7;250(1):120911–120915.
Journal cover image

Published In

J Phys Conf Ser

DOI

ISSN

1742-6588

Publication Date

December 7, 2010

Volume

250

Issue

1

Start / End Page

120911 / 120915

Location

England

Related Subject Headings

  • 51 Physical sciences
  • 0299 Other Physical Sciences
  • 0204 Condensed Matter Physics
  • 0202 Atomic, Molecular, Nuclear, Particle and Plasma Physics