DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM.
Publication
, Journal Article
Wolf, PD; Ideker, RE; Smith, WM
Published in: IEEE/Engineering in Medicine and Biology Society Annual Conference
December 1, 1986
A data stream formatter for a 128-channel mapping system was constructed which combines real-time variables and digitized waveforms into information packets. The variables are formatted by a microprocessor once per millisecond creating, with the digitized waveforms, a total throughput of 264 kB/s. Included in each ms of waveform data are four types of information: acquisition parameters, amplifier gain, short-term user variables, and external events. Synchronization of the header word generation with the analog sampling is done using control store to interrupt the microprocessor and control the data acquisition timing.
Duke Scholars
Published In
IEEE/Engineering in Medicine and Biology Society Annual Conference
Publication Date
December 1, 1986
Start / End Page
491 / 493
Citation
APA
Chicago
ICMJE
MLA
NLM
Wolf, P. D., Ideker, R. E., & Smith, W. M. (1986). DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM. IEEE/Engineering in Medicine and Biology Society Annual Conference, 491–493.
Wolf, P. D., R. E. Ideker, and W. M. Smith. “DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM.” IEEE/Engineering in Medicine and Biology Society Annual Conference, December 1, 1986, 491–93.
Wolf PD, Ideker RE, Smith WM. DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM. IEEE/Engineering in Medicine and Biology Society Annual Conference. 1986 Dec 1;491–3.
Wolf, P. D., et al. “DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM.” IEEE/Engineering in Medicine and Biology Society Annual Conference, Dec. 1986, pp. 491–93.
Wolf PD, Ideker RE, Smith WM. DATA STREAM FORMATTER FOR A CARDIAC MAPPING SYSTEM. IEEE/Engineering in Medicine and Biology Society Annual Conference. 1986 Dec 1;491–493.
Published In
IEEE/Engineering in Medicine and Biology Society Annual Conference
Publication Date
December 1, 1986
Start / End Page
491 / 493