Complications from prophylactic replacement of cardiac implantable electronic device generators in response to United States Food and Drug Administration recall: A systematic review and meta-analysis.
BACKGROUND: The number of cardiac implantable electronic device (CIED) recalls and advisories has increased over the past 3 decades, yet no consensus exists on how to best manage patients with these CIEDs, partially because rates of complications from prophylactic replacement are unknown. OBJECTIVE: The purpose of this study was to establish rates of complications when recalled CIED generators are replaced prophylactically. METHODS: We searched MEDLINE and the Cochrane Controlled Trials Register for reports of prophylactic replacement of recalled CIED generators. Studies with <20 subjects were excluded. We then conducted a meta-analysis of qualifying studies to determine the rates of combined major complications, mortality, and reoperation. RESULTS: We identified 7 citations that met our inclusion criteria and reported ≥1 end-points of interest. Four were single center, and 3 were multicenter. Six studies collected data retrospectively (n = 1213) and 1 prospectively (n = 222). Using a random effects model to combine data from all included studies, the rate of major complications was 2.5% (95% confidence interval [CI] 1.0%-4.5%). Combining data from 6 studies reporting mortality and reoperation, the rates were 0.5% (95% CI 0.1%-0.9%) and 2.5% (95% CI 0.8%-4.5%), respectively. CONCLUSION: Prophylactic replacement of recalled CIED generators is associated with a low mortality rate but nontrivial rates of other major complications similar to those reported when CIED generators are replaced for other reasons. Thus, when considering replacing a recalled CIED generator, known risks of elective generator replacement likely apply and can be weighed against risks associated with device failure.
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
DOI
EISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- United States Food and Drug Administration
- United States
- Risk Assessment
- Reoperation
- Postoperative Complications
- Medical Device Recalls
- Humans
- Equipment Failure
- Device Removal
- Defibrillators, Implantable
Citation
Published In
DOI
EISSN
Publication Date
Volume
Issue
Start / End Page
Location
Related Subject Headings
- United States Food and Drug Administration
- United States
- Risk Assessment
- Reoperation
- Postoperative Complications
- Medical Device Recalls
- Humans
- Equipment Failure
- Device Removal
- Defibrillators, Implantable