Skip to main content

Quiet transcranial magnetic stimulation: Status and future directions.

Publication ,  Conference
Peterchev, AV; Murphy, DLK; Goetz, SM
Published in: Annu Int Conf IEEE Eng Med Biol Soc
2015

A significant limitation of transcranial magnetic stimulation (TMS) is that the magnetic pulse delivery is associated with a loud clicking sound as high as 140 dB resulting from electromagnetic forces. The loud noise significantly impedes both basic research and clinical applications of TMS. It effectively makes TMS less focal since every click activates auditory cortex, brainstem, and other connected regions, synchronously with the magnetic pulse. The repetitive clicking sound can induce neuromodulation that can interfere with and confound the intended effects at the TMS target. As well, there are known concerns regarding blinding of TMS studies, hearing loss, induction of tinnitus, as well as tolerability. Addressing this need, we are developing a quiet TMS (qTMS) device that incorporates two key concepts: First, the dominant frequency components of the TMS pulse sound (typically 2-5 kHz) are shifted to higher frequencies that are above the human hearing upper threshold of about 20 kHz. Second, the TMS coil is designed electrically and mechanically to generate suprathreshold electric field pulses while minimizing the sound emitted at audible frequencies (<; 20 kHz). The enhanced acoustic properties of the coil are accomplished with a novel, layered coil design. We summarize a proof-of-concept qTMS prototype demonstrating noise loudness reduction by 19 dB(A) with ultrabrief pulses at conventional amplitudes. Further, we outline next steps to accomplish further sound reduction and suprathreshold pulse amplitudes.

Duke Scholars

Published In

Annu Int Conf IEEE Eng Med Biol Soc

DOI

EISSN

2694-0604

Publication Date

2015

Volume

2015

Start / End Page

226 / 229

Location

United States

Related Subject Headings

  • Transcranial Magnetic Stimulation
  • Tinnitus
  • Humans
  • Hearing
  • Equipment Design
  • Auditory Cortex
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Peterchev, A. V., Murphy, D. L. K., & Goetz, S. M. (2015). Quiet transcranial magnetic stimulation: Status and future directions. In Annu Int Conf IEEE Eng Med Biol Soc (Vol. 2015, pp. 226–229). United States. https://doi.org/10.1109/EMBC.2015.7318341
Peterchev, Angel V., David L. K. Murphy, and Stefan M. Goetz. “Quiet transcranial magnetic stimulation: Status and future directions.” In Annu Int Conf IEEE Eng Med Biol Soc, 2015:226–29, 2015. https://doi.org/10.1109/EMBC.2015.7318341.
Peterchev AV, Murphy DLK, Goetz SM. Quiet transcranial magnetic stimulation: Status and future directions. In: Annu Int Conf IEEE Eng Med Biol Soc. 2015. p. 226–9.
Peterchev, Angel V., et al. “Quiet transcranial magnetic stimulation: Status and future directions.Annu Int Conf IEEE Eng Med Biol Soc, vol. 2015, 2015, pp. 226–29. Pubmed, doi:10.1109/EMBC.2015.7318341.
Peterchev AV, Murphy DLK, Goetz SM. Quiet transcranial magnetic stimulation: Status and future directions. Annu Int Conf IEEE Eng Med Biol Soc. 2015. p. 226–229.

Published In

Annu Int Conf IEEE Eng Med Biol Soc

DOI

EISSN

2694-0604

Publication Date

2015

Volume

2015

Start / End Page

226 / 229

Location

United States

Related Subject Headings

  • Transcranial Magnetic Stimulation
  • Tinnitus
  • Humans
  • Hearing
  • Equipment Design
  • Auditory Cortex