Metal Artifact-Free and Dose-Efficient Computed Tomography Imaging of a Cardiac Prosthetic Valve Using Spectral Shaping of the X-ray Beam: A Case Report.

Publication ,  Journal Article
Schabel, C; Sekaran, NK; Wang, A; Hurwitz, L
Published in: J Comput Assist Tomogr
2018

Bioprosthetic valve (BPV) thrombosis can be challenging to diagnose with cardiac computed tomography owing to metal artifacts of the BPV. In this case report, an optimized metal artifact reduction protocol using a third-generation dual-source multidetector computed tomographic scanner with high kVp (Sn150kVp) and tin (Sn) filtration and high temporal resolution yielded high-quality, artifact-free, dynamic images of a thrombosed mitral BPV.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

J Comput Assist Tomogr

DOI

EISSN

1532-3145

Publication Date

2018

Volume

42

Issue

3

Start / End Page

467 / 468

Location

United States

Related Subject Headings

  • X-Rays
  • Tomography, X-Ray Computed
  • Signal-To-Noise Ratio
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Metals
  • Humans
  • Heart Valve Prosthesis
  • Female
  • Coronary Thrombosis
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Schabel, C., Sekaran, N. K., Wang, A., & Hurwitz, L. (2018). Metal Artifact-Free and Dose-Efficient Computed Tomography Imaging of a Cardiac Prosthetic Valve Using Spectral Shaping of the X-ray Beam: A Case Report. J Comput Assist Tomogr, 42(3), 467–468. https://doi.org/10.1097/RCT.0000000000000692
Schabel, Christoph, Nishant K. Sekaran, Andrew Wang, and Lynne Hurwitz. “Metal Artifact-Free and Dose-Efficient Computed Tomography Imaging of a Cardiac Prosthetic Valve Using Spectral Shaping of the X-ray Beam: A Case Report.J Comput Assist Tomogr 42, no. 3 (2018): 467–68. https://doi.org/10.1097/RCT.0000000000000692.
Schabel, Christoph, et al. “Metal Artifact-Free and Dose-Efficient Computed Tomography Imaging of a Cardiac Prosthetic Valve Using Spectral Shaping of the X-ray Beam: A Case Report.J Comput Assist Tomogr, vol. 42, no. 3, 2018, pp. 467–68. Pubmed, doi:10.1097/RCT.0000000000000692.

Published In

J Comput Assist Tomogr

DOI

EISSN

1532-3145

Publication Date

2018

Volume

42

Issue

3

Start / End Page

467 / 468

Location

United States

Related Subject Headings

  • X-Rays
  • Tomography, X-Ray Computed
  • Signal-To-Noise Ratio
  • Radiation Dosage
  • Nuclear Medicine & Medical Imaging
  • Metals
  • Humans
  • Heart Valve Prosthesis
  • Female
  • Coronary Thrombosis