Liver fat quantification in photon counting CT in head to head comparison with clinical MRI - First experience.
PURPOSE: To compare liver fat quantification between MRI and photon-counting CT (PCCT). METHOD: A cylindrical phantom with inserts containing six concentrations of oil (0, 10, 20, 30, 50 and 100%) and oil-iodine mixtures (0, 10, 20, 30 and 50% fat +3 mg/mL iodine) was imaged with a PCCT (NAEOTOM Alpha) and a 1.5 T MRI system (MR 450w, IDEAL-IQ sequence), using clinical parameters. An IRB-approved prospective clinical evaluation included 12 obese adult patients with known fatty liver disease (seven women, mean age: 61.5 ± 13 years, mean BMI: 30.3 ± 4.7 kg/m2). Patients underwent a same-day clinical MRI and PCCT of the abdomen. Liver fat fractions were calculated for four segments (I, II, IVa and VII) using in- and opposed-phase on MRI ((Meanin - Meanopp)/2*Meanin) and iodine-fat, tissue decomposition analysis in PCCT (Syngo.Via VB60A). CT and MRI Fat fractions were compared using two-sample t-tests with equal variance. Statistical analysis was performed using RStudio (Version1.4.1717). RESULTS: Phantom results showed no significant differences between the known fat fractions (P = 0.32) or iodine (P = 0.6) in comparison to PCCT-measured concentrations, and no statistically significant difference between known and MRI-measured fat fractions (P = 0.363). In patients, the mean fat signal fraction measured on MRI and PCCT was 13.1 ± 9.9% and 12.0 ± 9.0%, respectively, with an average difference of 1.1 ± 1.9% between the modalities (P = 0.138). CONCLUSION: First experience shows promising accuracy of liver fat fraction quantification for PCCT in obese patients. This method may improve opportunistic screening for CT in the future.
Duke Scholars
Altmetric Attention Stats
Dimensions Citation Stats
Published In
DOI
EISSN
Publication Date
Volume
Start / End Page
Location
Related Subject Headings
- Tomography, X-Ray Computed
- Reproducibility of Results
- Nuclear Medicine & Medical Imaging
- Middle Aged
- Male
- Magnetic Resonance Imaging
- Liver
- Humans
- Female
- Fatty Liver
Citation
Published In
DOI
EISSN
Publication Date
Volume
Start / End Page
Location
Related Subject Headings
- Tomography, X-Ray Computed
- Reproducibility of Results
- Nuclear Medicine & Medical Imaging
- Middle Aged
- Male
- Magnetic Resonance Imaging
- Liver
- Humans
- Female
- Fatty Liver